  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lewisville",
      "address_1": "3451 Plano Pkwy",
      "reason_for_recall": "There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.",
      "address_2": "",
      "product_quantity": "36 units",
      "code_info": "All lot codes",
      "center_classification_date": "20150622",
      "distribution_pattern": "Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.",
      "state": "TX",
      "product_description": "ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.",
      "report_date": "20150701",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Orthofix, Inc",
      "recall_number": "Z-1819-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71262",
      "termination_date": "20160217",
      "more_code_info": "",
      "recall_initiation_date": "20150511",
      "postal_code": "75056-9453",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}