  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Milwaukee",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85297",
      "recalling_firm": "WELCH ALLYN, INC/MORTARA",
      "address_1": "7865 N 86th St",
      "address_2": "N/A",
      "postal_code": "53224-3431",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.",
      "recall_number": "Z-1812-2020",
      "product_description": "H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP.     The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.",
      "product_quantity": "78,521 units total",
      "reason_for_recall": "Impacted Welch Allyn products may not meet the  Defibrillation Withstand  requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.",
      "recall_initiation_date": "20200318",
      "center_classification_date": "20200429",
      "termination_date": "20260209",
      "report_date": "20200506",
      "code_info": "Model Numbers: H12PLUS-XXX-XXXXX, 9293-017-50, 9293-017-51, 9293-026-50, 9293-026-51."
    }
  ]
}