      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87962",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.",
      "recall_number": "Z-1796-2021",
      "product_description": "Ingenia Ambition S - Model Number: 781359  Diagnostic MR system.",
      "product_quantity": "156 MR systems",
      "reason_for_recall": "The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.",
      "recall_initiation_date": "20210423",
      "center_classification_date": "20210604",
      "report_date": "20210616",
      "code_info": "REF 781359/ UDI-DI 00884838090057 Serial numbers:  48000 48001 48003 48004 48009 48005 48015 48017 48020 48011 48022 48012 48014 48028 48026 48037 48023 48043 48019 48160 48144 48180 48112 48052  ***Updated November 2023*** Added Serial Numbers  48051 48010 48038 48018 48039 48027 48122 48046 48036 48047 48133 48056 48169 48148 48154 48055 48128 48194 48173 48134 48040 48155 48186 48035 48181 48188 48198"
    }
  ]
}