{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Hennigsdorf",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98561",
      "recalling_firm": "Brahms GmbH",
      "address_1": "Neuendorfstr. 25",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide distribution. US states: NC, TX, MN, UT, NY, FL, OH, IL, CT, CA, GA.  Other countries: DE, AT, AL, AU, BG, BR, CA, CH, CZ, DK, ET, ES, FR, GB, GR, HK, HU, ID, IL, IT, KW, KZ, LT, MK, ML, MX, NL, NO, PA, PL, PT, RO, SE, SI, TH, TU, TW, UA, ZA",
      "recall_number": "Z-1795-2026",
      "product_description": "B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N.  in vitro diagnostic test for Placental Growth Factor",
      "product_quantity": "5,254",
      "reason_for_recall": "Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve.  If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.",
      "recall_initiation_date": "20260312",
      "center_classification_date": "20260408",
      "report_date": "20260415",
      "code_info": "UDI: 04260157632196. Lot(Expiration): 59172(2026-11-23), 859075N(2026-11-23)",
      "more_code_info": ""
    }
  ]
}