  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Durham",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84737",
      "recalling_firm": "bioMerieux, Inc.",
      "address_1": "100 Rodolphe St",
      "address_2": "N/A",
      "postal_code": "27712-9402",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide including AU, BE, CH, DE, ES, FR, IT, HK/CN; MX, NL, RU, SE/FI, ZA.",
      "recall_number": "Z-1793-2020",
      "product_description": "PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units),  P1028888-001 PSU (used in Zebra LP2824 Plus printers)     PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).",
      "product_quantity": "N/A",
      "reason_for_recall": "The recall has been initiated due to potential for overheating or  a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.",
      "recall_initiation_date": "20200124",
      "center_classification_date": "20200423",
      "termination_date": "20230111",
      "report_date": "20200429",
      "code_info": "Instrument Serial Number   AS180-00028 1035196   AS180-00067 1007580   AS180-00084 2055833   AS180-00105 2025461   AS180-00138 2025461   AS180-00179 1039938   AS180-00217 1033454   AS180-00248 1029262   AS180-00278 1031886   AS180-00300 1033400   AS180-00306 1034034   AS180-00326 1034190   AS180-00336 1041241   AS180-00339 2054793   AS180-00357 2055775   AS180-00381 1073849   AS180-00385 1054234   AS180-00398 1052053   AS180-00402 2055620   AS180-00405 1063739   AS180-00418 1042188   AS180-00420 1040343   AS180-00428 1030890   AS180-00453 1031472   AS180-00468 1043640   AS180-00476 1029272   AS180-00478 1053432   AS180-00479 1011691   AS180-00480 5001554   AS180-00481 1039588   AS180-00482 1034363   AS180-00483 1029990   AS180-00484 1049298   AS180-00485 AU01   AS180-00486 1047965   AS180-00487 1074074   AS180-00488 1000283   AS180-00489 1035594   AS180-00490 1047965   AS180-00491 1035549   AS180-00492 1070878   AS180-00493 1090075   AS180-00494 1039290   AS180-00495 1029148   AS180-00500 1007884   AS180-00509 1065153"
    }
  ]
}