{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Santa Clarita",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94381",
      "recalling_firm": "TriMed Inc.",
      "address_1": "27533 Avenue Hopkins",
      "address_2": "N/A",
      "postal_code": "91355-3910",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, \tOH, OK, PA, TN, TX, UT, VA, WA, WI.",
      "recall_number": "Z-1791-2024",
      "product_description": "TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only  TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only",
      "product_quantity": "142 units",
      "reason_for_recall": "Nail system contains an out of specification driver-socket which  could result in inability to pick up nails or apply sufficient torque.",
      "recall_initiation_date": "20240319",
      "center_classification_date": "20240510",
      "report_date": "20240522",
      "code_info": "UDI: 00842188127878/ Lot # QM23086"
    }
  ]
}