  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marburg",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85282",
      "recalling_firm": "CSL Behring GmbH",
      "address_1": "Emil-Von-Behring-Str. 76",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "IN, KS, MO, OH, VA  Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China  Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary  India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain  Sweden, Switzerland, Taiwan",
      "recall_number": "Z-1790-2020",
      "product_description": "Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum.  Siemens Material Number (SMN):  10445997",
      "product_quantity": "297 PCK US; 1247 PCK OUS",
      "reason_for_recall": "Negative bias for the carbohydrate-deficient transferrin (CDT) measurement  in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring",
      "recall_initiation_date": "20200302",
      "center_classification_date": "20200423",
      "termination_date": "20201005",
      "report_date": "20200429",
      "code_info": "US Lot number: 49822   UDI:  842768018534  OUS: Lot numbers: 49729, 49849, 49985, 50084    Exp. Date: 2020-04-11"
    }
  ]
}