  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96674",
      "recalling_firm": "Echopixel, Inc.",
      "address_1": "780 Montague Expy Ste 305",
      "address_2": "N/A",
      "postal_code": "95131-1317",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of CA, OH, NJ, PA, MA, FL, CT, TX, LA, MI, MN and the country of CA.",
      "recall_number": "Z-1785-2025",
      "product_description": "True 3D Viewer, Model: EP-00003",
      "product_quantity": "18",
      "reason_for_recall": "With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from CT/MRI/ultrasound, and pre-operative analysis of surgical options), can result in incorrect display of image orientation, which could result in misdiagnosis or incorrect surgical treatment plan.",
      "recall_initiation_date": "20180221",
      "center_classification_date": "20250516",
      "report_date": "20250528",
      "code_info": "UDI-DI: 00851325007003.  Full release software versions: 1.6.1.3080.428, released 13-Apr-2017 1.6.2.3670.435, released 07-Dec-2017  Develop Versions, for research use: 0.00.2695.422, released 15-Feb-2017 0.00.2927.423, released 15-Feb-2017 0.00.2964.423, released 24-Feb-2017 0.00.3102.428, released 04-Apr-2017 0.00.3245.428, released 05-May-2017 0.00.3302.431, released 02-Jun-2017 0.00.3344.432, released 18-Jul-2017 0.00.3412.433, released 14-Aug-2017 0.00.3656.442, released 09-Nov-2017 0.00.3669.442, released 05-Dec-2017 0.00.3683.445, released 02-Jan-2018 0.00.3702.447, released 11-Jan-2018 0.00.3730.447, released 08-Feb-2018"
    }
  ]
}