  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90777",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of WV, NC.",
      "recall_number": "Z-1772-2022",
      "product_description": "The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.",
      "product_quantity": "2 units",
      "reason_for_recall": "***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI.  Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg",
      "recall_initiation_date": "20220801",
      "center_classification_date": "20220919",
      "termination_date": "20241213",
      "report_date": "20220928",
      "code_info": "GTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023"
    }
  ]
}