      "status": "Terminated",
      "city": "Hertford",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65630",
      "recalling_firm": "Del Mar Reynolds Medical, Ltd.",
      "address_1": "1-2 Harforde Ct., Foxholes",
      "address_2": "Business Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including South Carolina and internationally to Canada and Great Britian.",
      "recall_number": "Z-1771-2013",
      "product_description": "Spacelabs ARKON Anesthesia Workstation.     Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer.",
      "product_quantity": "9 units (4 in the US and 5 outside US)",
      "reason_for_recall": "Spacelabs ARKON Anesthesia System is recalled due to software defect.  The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.",
      "recall_initiation_date": "20130607",
      "center_classification_date": "20130719",
      "termination_date": "20130829",
      "report_date": "20130731",
      "code_info": "ARKN-000001, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000016, ARKN-000017, ARKN-000018, ARKN-000020, ARKN-000021.",
      "more_code_info": ""
    }
  ]
}