  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82878",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Rd",
      "address_2": "N/A",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US nationwide, including CA, CT, IL, IN, KY, MI, MN, MO, MT, NM, NY, OH, and WI. Foreign distribution to Canada, Mexico, and South Korea.",
      "recall_number": "Z-1769-2019",
      "product_description": "Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588.  Used in cardio-pulmonary bypass.",
      "product_quantity": "72",
      "reason_for_recall": "An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.",
      "recall_initiation_date": "20181017",
      "center_classification_date": "20190611",
      "termination_date": "20200430",
      "report_date": "20190619",
      "code_info": "01338  01339  01340  01341  01342  01349  01350  01351  01352  01353  01354  01355  01356  01357  01358  01360  01361  01362  01363  01365  01366  01367  01369  01370  01371  01372  01373  01374  01375  01376  01377  01378  01379  Lot/Serial  01380  01381  01382  01383  01384  01385  01386  01387  01388  01389  01394  01395  01396  01397  01398  01405  01406  01407  01408  01410  01411  01413  01414  01415  01416  01418  01420  01421  01422  01423  01424  01425  01426  Lot/Serial  01427  01428  01429  01430  01431  01432  01434  01435  1718  2317  4417  9999    UDI 00886799000588"
    }
  ]
}