  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76723",
      "recalling_firm": "Nihon Kohden America Inc",
      "address_1": "15353 Barranca Pkwy",
      "address_2": "N/A",
      "postal_code": "92618-2216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide Distribution including the countries of Germany, Spain, Brazil, Paraguay, Bolivia, Chile, Colombia, Costa Rica, Mexico, Peru, Pakistan, Israel, Qatar, Saudi Arabia, South Africa, Tunisia, United Arab Emirates, Korea, Malaysia, Singapore, Thailand",
      "recall_number": "Z-1768-2017",
      "product_description": "Bedside monitor  CSM-1901(Life Scope G9) with main unit CU-192RA.  The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).",
      "product_quantity": "654 units total (230 units in US)",
      "reason_for_recall": "The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.",
      "recall_initiation_date": "20170313",
      "center_classification_date": "20170410",
      "termination_date": "20171001",
      "report_date": "20170419",
      "code_info": "Lot/Unit Codes: G9  Serial Number: 230 (UDI: 01049319211121682100230)  Serial Number: 232 (UDI: 01049319211121682100232)    Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software.    Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower.",
      "more_code_info": ""
    }
  ]
}