  "results": [
    {
      "status": "Ongoing",
      "city": "Queensbury",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94349",
      "recalling_firm": "Angiodynamics, Inc.",
      "address_1": "603 Queensbury Ave",
      "address_2": "N/A",
      "postal_code": "12804-7619",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "VA",
      "recall_number": "Z-1767-2024",
      "product_description": "DURAMAX STACKED TIP 24CM STR. BASIC KIT, 5-UP PG-\tIndicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis.    Product Number: H787103028035",
      "product_quantity": "555 units",
      "reason_for_recall": "Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.",
      "recall_initiation_date": "20240318",
      "center_classification_date": "20240503",
      "report_date": "20240515",
      "code_info": "UDI-DI: 25051684011836   Lot Numbers: 5807194, 5809616, 5809462"
    }
  ]
}