  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Smithfield",
      "state": "RI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90762",
      "recalling_firm": "Alcor Scientific, Inc.",
      "address_1": "20 Thurber Blvd",
      "address_2": "",
      "postal_code": "02917-1858",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea",
      "recall_number": "Z-1751-2022",
      "product_description": "iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the  iSED Automated ESR Analyzer   PN: 112-13-002",
      "product_quantity": "38 units",
      "reason_for_recall": "Primary Pump Tubing Spare Part is incorrect.  If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSED\u0019s Probe Tube",
      "recall_initiation_date": "20220720",
      "center_classification_date": "20220915",
      "report_date": "20220921",
      "code_info": "Lot 4/22",
      "more_code_info": ""
    }
  ]
}