  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Farnborough",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96719",
      "recalling_firm": "Remote Diagnostic Technologies Ltd.",
      "address_1": "Farnborough Aerospace Centre",
      "address_2": "Aerospace Boulevard",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.",
      "recall_number": "Z-1729-2025",
      "product_description": "Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R,  00-1026-R, 00-1004, 00-1007",
      "product_quantity": "6961",
      "reason_for_recall": "Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.",
      "recall_initiation_date": "20250415",
      "center_classification_date": "20250507",
      "report_date": "20250514",
      "code_info": "Software versions prior to vx.40.  REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655",
      "more_code_info": ""
    }
  ]
}