{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mounds View",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92328",
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "address_1": "8200 Coral Sea St Ne",
      "address_2": "",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - Worldwide Distribution",
      "recall_number": "Z-1728-2023",
      "product_description": "ICD CROME DR MRI IS1 DF4, Model Number DDPC3D4;  Implantable Cardioverter Defibrillators",
      "product_quantity": "2018 Units",
      "reason_for_recall": "There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.",
      "recall_initiation_date": "20230510",
      "center_classification_date": "20230616",
      "report_date": "20230628",
      "code_info": "GTIN 00763000178437, Lot Serial Numbers:  RSQ600736S, RSQ600745S, RSQ600905S, RSQ600975S, RSQ601406S, RSQ600113S, RSQ600114S, RSQ600115S, RSQ600255S, RSQ600256S, RSQ600257S, RSQ600258S, RSQ600261S, RSQ600740S, RSQ601122S, RSQ601137S, RSQ601177S, RSQ601178S, RSQ601179S, RSQ601181S, RSQ601227S, RSQ601228S, RSQ601229S, RSQ601375S, RSQ601377S, RSQ600055S, RSQ600163S, RSQ600164S, RSQ600936S, RSQ601151S, RSQ600049S, RSQ600051S, RSQ600713S, RSQ600991S, RSQ601411S, RSQ601416S, RSQ600218S, RSQ600237S, RSQ600679S, RSQ600712S, RSQ600723S, RSQ600772S, RSQ600074S, RSQ600090S, RSQ600012S, RSQ600053S, RSQ600811S, RSQ600843S, RSQ600912S, RSQ600913S, RSQ600195S, RSQ600894S, RSQ600981S, RSQ600748S, RSQ601201S, RSQ601378S, RSQ601479S, RSQ601490S, RSQ600782S, RSQ600665S, RSQ600091S, RSQ600105S, RSQ600106S, RSQ600130S, RSQ600174S, RSQ600183S, RSQ600185S, RSQ600216S, RSQ600767S, RSQ600788S, RSQ600799S, RSQ601445S, RSQ601447S, RSQ601448S, RSQ601451S, RSQ601452S, RSQ600704S, RSQ600743S, RSQ600773S, RSQ600808S, RSQ600262S, RSQ602018S, RSQ600101S, RSQ600738S, RSQ600093S, RSQ600125S, RSQ600128S, RSQ600140S, RSQ600678S, RSQ600694S, RSQ600737S, RSQ600744S, RSQ600750S, RSQ600753S, RSQ600754S, RSQ600780S, RSQ600789S, RSQ600807S, RSQ600829S, RSQ600906S, RSQ600931S, RSQ601212S, RSQ601413S, RSQ600097S, RSQ600232S, RSQ600233S, RSQ600247S, RSQ600705S, RSQ600719S, RSQ600721S, RSQ600751S, RSQ600752S, RSQ600812S, RSQ600821S, RSQ600858S, RSQ600859S, RSQ600896S, RSQ600897S, RSQ600915S, RSQ600939S, RSQ600986S, RSQ600989S, RSQ601078S, RSQ601193S, RSQ601194S, RSQ601203S, RSQ601205S, RSQ600790S, RSQ600787S, RSQ600938S, RSQ600842S, RSQ600071S, RSQ600078S, RSQ600191S, RSQ600225S, RSQ600231S, RSQ600742S, RSQ600759S, RSQ601010S, RSQ601216S, RSQ601395S, RSQ601430S, RSQ600786S, RSQ600098S, RSQ600135S, RSQ600252S, RSQ600080S, RSQ600085S, RSQ601189S, RSQ600700S, RSQ600677S, RSQ602013S, RSQ602014S, RSQ600082S, RSQ600133S, RSQ600136S, RSQ600137S, RSQ600152S, RSQ600166S, RSQ600224S, RSQ600739S, RSQ600941S, RSQ601169S, RSQ601154S, RSQ601155S, RSQ600054S, RSQ600058S, RSQ600067S, RSQ600076S, RSQ600131S, RSQ600143S, RSQ600148S, RSQ600184S, RSQ600710S, RSQ600934S, RSQ600958S, RSQ600971S, RSQ601183S, RSQ600718S, RSQ600760S, RSQ601196S, RSQ600695S, RSQ600075S, RSQ601923S, RSQ600854S, RSQ601211S, RSQ601446S, RSQ600746S, RSQ602012S, RSQ600702S, RSQ600703S, RSQ600709S, RSQ600922S, RSQ600978S, RSQ601152S, RSQ601210S, RSQ601382S, RSQ601393S, RSQ601443S, RSQ601450S, RSQ600918S, RSQ600948S, RSQ600102S, RSQ600155S, RSQ600160S, RSQ600701S, RSQ600763S, RSQ600940S, RSQ600815S, RSQ600162S, RSQ600177S, RSQ600178S, RSQ600179S, RSQ600181S, RSQ600209S, RSQ600213S, RSQ600222S, RSQ600227S, RSQ600228S, RSQ600240S, RSQ600610S, RSQ600641S, RSQ600657S, RSQ600683S, RSQ600693S, RSQ600698S, RSQ600733S, RSQ600734S, RSQ600741S, RSQ600749S, RSQ600758S, RSQ600761S, RSQ600762S, RSQ600764S, RSQ600765S, RSQ600766S, RSQ600775S, RSQ600783S, RSQ600817S, RSQ600830S, RSQ600908S, RSQ600920S, RSQ600943S, RSQ600966S, RSQ600968S, RSQ600974S, RSQ600979S, RSQ601006S, RSQ601170S, RSQ601174S, RSQ601175S, RSQ601176S, RSQ601218S, RSQ601403S, RSQ601410S, RSQ601435S, RSQ601438S, RSQ601439S, RSQ601440S, RSQ601441S, RSQ601976S, RSQ600935S, RSQ601157S, RSQ600860S, RSQ600890S, RSQ601618S, RSQ600118S, RSQ600068S, RSQ600081S, RSQ600779S, RSQ601592S, RSQ600169S, RSQ600716S, RSQ600717S, RSQ600769S, RSQ600837S, RSQ600923S, RSQ600930S, RSQ600944S, RSQ600988S, RSQ600992S, RSQ600995S, RSQ601110S, RSQ601132S, RSQ601159S, RSQ601167S, RSQ601182S, RSQ601188S, RSQ601191S, RSQ601224S, RSQ601226S, RSQ600785S, RSQ600793S, RSQ600803S, RSQ600804S, RSQ600816S, RSQ600824S, RSQ600828S, RSQ600899S, RSQ600903S, RSQ600904S, RSQ600973S, RSQ600983S, RSQ600985S, RSQ601002S, RSQ601148S, RSQ601156S, RSQ601162S, RSQ601163S, RSQ601164S, RSQ601165S, RSQ601166S, RSQ600919S, RSQ601079S, RSQ601862S, RSQ601921S, RSQ602278S, RSQ602285S, RSQ602287S, RSQ600089S, RSQ600798S, RSQ600802S, RSQ600996S, RSQ600998S, RSQ601503S, RSQ601536S, RSQ601550S, RSQ601554S, RSQ601555S, RSQ601582S, RSQ601584S, RSQ601585S, RSQ601586S, RSQ601588S, RSQ601589S, RSQ601593S, RSQ601597S, RSQ601600S, RSQ601601S, RSQ601602S, RSQ601603S, RSQ601604S, RSQ601605S, RSQ601606S, RSQ601607S, RSQ601609S, RSQ601610S, RSQ601615S, RSQ601617S, RSQ601619S, RSQ601627S, RSQ601628S, RSQ601629S, RSQ601630S, RSQ601632S, RSQ601634S, RSQ601635S, RSQ601636S, RSQ601639S, RSQ601641S, RSQ601643S, RSQ601644S, RSQ601646S, RSQ601647S, RSQ601648S, RSQ601649S, RSQ601650S, RSQ601651S, RSQ601652S, RSQ601653S, RSQ601661S, RSQ601662S, RSQ601663S, RSQ601664S, RSQ601667S, RSQ601670S, RSQ601671S, RSQ601673S, RSQ601674S, RSQ601675S, RSQ601677S, RSQ601679S, RSQ601680S, RSQ601682S, RSQ601684S, RSQ601686S, RSQ601687S, RSQ601689S, RSQ601690S, RSQ601691S, RSQ601693S, RSQ601694S, RSQ601695S, RSQ601696S, RSQ601697S, RSQ601698S, RSQ601699S, RSQ601871S, RSQ601872S, RSQ601874S, RSQ601876S, RSQ601883S, RSQ601892S, RSQ601902S, RSQ601907S, RSQ601909S, RSQ601924S, RSQ601933S, RSQ601935S, RSQ601936S, RSQ601937S, RSQ601938S, RSQ601942S, RSQ601952S, RSQ601957S, RSQ602001S, RSQ602005S, RSQ602007S, RSQ602229S, RSQ602289S, RSQ602290S, RSQ602294S, RSQ602295S, RSQ602296S, RSQ602297S, RSQ602298S, RSQ602299S, RSQ602303S, RSQ602304S, RSQ602308S, RSQ602310S, RSQ602311S, RSQ602395S, RSQ602397S, RSQ602398S, RSQ602399S, RSQ602400S, RSQ602404S, RSQ600263S, RSQ601103S, RSQ600070S, RSQ600153S, RSQ600167S, RSQ600175S, RSQ600778S, RSQ600791S, RSQ600910S, RSQ600957S, RSQ600801S, RSQ600670S, RSQ600747S, RSQ601153S, RSQ600057S, RSQ600990S, RSQ600776S, RSQ600831S, RSQ600202S, RSQ600204S, RSQ600266S, RSQ600079S, RSQ600666S, RSQ601433S, RSQ600197S, RSQ600928S, RSQ601161S, RSQ601214S, RSQ601219S, RSQ601220S, RSQ601476S, RSQ600159S, RSQ600129S, RSQ600146S, RSQ600147S, RSQ600210S, RSQ600056S, RSQ600094S, RSQ600212S, RSQ600675S, RSQ600728S, RSQ601421S, RSQ601434S, RSQ600120S, RSQ600123S, RSQ600607S, RSQ600668S, RSQ600691S, RSQ600768S, RSQ600774S, RSQ600927S, RSQ600929S, RSQ600959S, RSQ600960S, RSQ600987S, RSQ601001S, RSQ600065S, RSQ600122S, RSQ600127S, RSQ600154S, RSQ600156S, RSQ600796S, RSQ600813S, RSQ600836S, RSQ600236S, RSQ600241S, RSQ601108S, RSQ600194S, RSQ600246S, RSQ600460S, RSQ600633S, RSQ601150S, RSQ600108S, RSQ600221S, RSQ600254S, RSQ600781S, RSQ600823S, RSQ601160S, RSQ601172S, RSQ601442S, RSQ600112S, RSQ600244S, RSQ600248S, RSQ600171S, RSQ600207S, RSQ600259S, RSQ601168S, RSQ600066S, RSQ600083S, RSQ600911S, RSQ600648S, RSQ600794S, RSQ601147S, RSQ600243S, RSQ601190S, RSQ600141S, RSQ600809S, RSQ600061S, RSQ600073S, RSQ600220S, RSQ600253S, RSQ600895S, RSQ600917S, RSQ600970S, RSQ600993S, RSQ601000S, RSQ600157S, RSQ600186S, RSQ600201S, RSQ600226S, RSQ600726S, RSQ600771S, RSQ600900S, RSQ600907S, RSQ600951S, RSQ600982S, RSQ600984S, RSQ600165S, RSQ601186S, RSQ600921S, RSQ600924S, RSQ601474S, RSQ600205S, RSQ600111S, RSQ600092S, RSQ601134S, RSQ601138S, RSQ600139S, RSQ600084S, RSQ600800S, RSQ600059S, RSQ600110S, RSQ600119S, RSQ600149S, RSQ600150S, RSQ601668S, RSQ601910S, RSQ601919S;  GTIN 00763000178444, Lot Serial Numbers:  RSQ601381S, RSQ601864S, RSQ601577S, RSQ601725S, RSQ602374S, RSQ602588S, RSQ600669S, RSQ601974S, RSQ602244S, RSQ602281S, RSQ602350S, RSQ602191S, RSQ602565S, RSQ602566S, RSQ602118S, RSQ602139S, RSQ601767S, RSQ602490S, RSQ601983S, RSQ601745S, RSQ602226S, RSQ602536S, RSQ602320S, RSQ602321S, RSQ602370S, RSQ601884S, RSQ601925S, RSQ601987S, RSQ602234S, RSQ602235S, RSQ602271S, RSQ602280S, RSQ600290S, RSQ600459S, RSQ600435S, RSQ601093S, RSQ600582S, RSQ601380S, RSQ600593S, RSQ600024S, RSQ600489S, RSQ600494S, RSQ600554S, RSQ601029S, RSQ601032S, RSQ601394S, RSQ601702S, RSQ600047S, RSQ600027S, RSQ600031S, RSQ600042S, RSQ600585S, RSQ600602S, RSQ601042S, RSQ601396S, RSQ601414S, RSQ601437S, RSQ601580S, RSQ601621S, RSQ601880S, RSQ601903S, RSQ601912S, RSQ602367S, RSQ602501S, RSQ601456S, RSQ600977S, RSQ602349S, RSQ601866S, RSQ601918S, RSQ602543S, RSQ602277S, RSQ602343S, RSQ600478S, RSQ601007S, RSQ601386S, RSQ601547S, RSQ602272S, RSQ602306S, RSQ600653S, RSQ601288S, RSQ602193S, RSQ602273S, RSQ601081S, RSQ600550S, RSQ600933S, RSQ601701S, RSQ602130S, RSQ600926S, RSQ601703S, RSQ601085S, RSQ601231S, RSQ601793S, RSQ600618S, RSQ600620S, RSQ600655S, RSQ600937S, RSQ600946S, RSQ601265S, RSQ601737S, RSQ600591S, RSQ601383S, RSQ601424S, RSQ600650S, RSQ600581S, RSQ600640S, RSQ600643S, RSQ600647S, RSQ600708S, RSQ600832S, RSQ600845S, RSQ600945S, RSQ601009S, RSQ601077S, RSQ601199S, RSQ601260S, RSQ601264S, RSQ601419S, RSQ601454S, RSQ601573S, RSQ601812S, RSQ602208S, RSQ600855S, RSQ601384S, RSQ600605S, RSQ601289S, RSQ601855S, RSQ600530S, RSQ601091S, RSQ600652S, RSQ600032S, RSQ600046S, RSQ600395S, RSQ600635S, RSQ600725S, RSQ600727S, RSQ600806S, RSQ600557S, RSQ601404S, RSQ602225S, RSQ600560S, RSQ600562S, RSQ600637S, RSQ600847S, RSQ601004S, RSQ601282S, RSQ601299S, RSQ601704S, RSQ601754S, RSQ602206S, RSQ600528S, RSQ602233S, RSQ601041S, RSQ601046S, RSQ601048S, RSQ601851S, RSQ602318S, RSQ602326S, RSQ602561S, RSQ600190S, RSQ601198S, RSQ601401S, RSQ602564S, RSQ600849S, RSQ601058S, RSQ601455S, RSQ601658S, RSQ602356S, RSQ600022S, RSQ600841S, RSQ600851S, RSQ600626S, RSQ601795S, RSQ601841S, RSQ602151S, RSQ600294S, RSQ601011S, RSQ601057S, RSQ601127S, RSQ601425S, RSQ600825S, RSQ600953S, RSQ601055S, RSQ601390S, RSQ600439S, RSQ600472S, RSQ600687S, RSQ600792S, RSQ601053S, RSQ601391S, RSQ601040S, RSQ601272S, RSQ601574S, RSQ601021S, RSQ601082S, RSQ600443S, RSQ600444S, RSQ600445S, RSQ600446S, RSQ600449S, RSQ600437S, RSQ600639S, RSQ600805S, RSQ601734S, RSQ601806S, RSQ601846S, RSQ601854S, RSQ601901S, RSQ600438S, RSQ600295S, RSQ600676S, RSQ601052S, RSQ601420S, RSQ601061S, RSQ601043S, RSQ601457S, RSQ601458S, RSQ601757S, RSQ601758S, RSQ600833S, RSQ601733S, RSQ601067S, RSQ600464S, RSQ600465S, RSQ600561S, RSQ600592S, RSQ601073S, RSQ601811S, RSQ600622S, RSQ600722S, RSQ600531S, RSQ601752S, RSQ601860S, RSQ602221S, RSQ600575S, RSQ600601S, RSQ600835S, RSQ600955S, RSQ601024S, RSQ601415S, RSQ601660S, RSQ601747S, RSQ600462S, RSQ600485S, RSQ600552S, RSQ600553S, RSQ600577S, RSQ600674S, RSQ600878S, RSQ601016S, RSQ601107S, RSQ601114S, RSQ601266S, RSQ601575S, RSQ601591S, RSQ601709S, RSQ601753S, RSQ602366S, RSQ600662S, RSQ600663S, RSQ600863S, RSQ600866S, RSQ600942S, RSQ600962S, RSQ600976S, RSQ600997S, RSQ600690S, RSQ600784S, RSQ600287S, RSQ600873S, RSQ601017S, RSQ601018S, RSQ601062S, RSQ601036S, RSQ601666S, RSQ601277S, RSQ601297S, RSQ601567S, RSQ600281S, RSQ601235S, RSQ601741S, RSQ601742S, RSQ601898S, RSQ602276S, RSQ600567S, RSQ600586S, RSQ601669S, RSQ601672S, RSQ601823S, RSQ601826S, RSQ602223S, RSQ602224S, RSQ602352S, RSQ602353S, RSQ602526S, RSQ601202S, RSQ601232S, RSQ601766S, RSQ601897S, RSQ602190S, RSQ602305S, RSQ602560S, RSQ601858S, RSQ602000S, RSQ602006S, RSQ602207S, RSQ602315S, RSQ602325S, RSQ602336S, RSQ602337S, RSQ602373S, RSQ602572S, RSQ602468S, RSQ601870S, RSQ600381S, RSQ600382S, RSQ600628S, RSQ600573S, RSQ600604S, RSQ600664S, RSQ600671S, RSQ600297S, RSQ600834S, RSQ600885S, RSQ600914S, RSQ600932S, RSQ600954S, RSQ600964S, RSQ600965S, RSQ601080S, RSQ601859S, RSQ601886S, RSQ601895S, RSQ602150S, RSQ602216S, RSQ602217S, RSQ602381S, RSQ602386S, RSQ602464S, RSQ602466S, RSQ600466S, RSQ600467S, RSQ600686S, RSQ600688S, RSQ601088S, RSQ601422S, RSQ601596S, RSQ601598S, RSQ601608S, RSQ601612S, RSQ601616S, RSQ602617S, RSQ602618S, RSQ601553S, RSQ601563S, RSQ601564S, RSQ601565S, RSQ601566S, RSQ600532S, RSQ600535S, RSQ600608S, RSQ601853S, RSQ601997S, RSQ602214S, RSQ602327S, RSQ602382S, RSQ602599S, RSQ602600S, RSQ600570S, RSQ601837S, RSQ601839S, RSQ600453S, RSQ601569S, RSQ601578S, RSQ601908S, RSQ601989S, RSQ602171S, RSQ602194S, RSQ602211S, RSQ602219S, RSQ602516S, RSQ602169S, RSQ602351S, RSQ602383S, RSQ602593S, RSQ602595S, RSQ601850S, RSQ601988S, RSQ602176S, RSQ602324S, RSQ600569S, RSQ600949S, RSQ601086S, RSQ601090S, RSQ601453S, RSQ601546S, RSQ601583S, RSQ601827S, RSQ601879S, RSQ601999S, RSQ602015S, RSQ602021S, RSQ602142S, RSQ602192S, RSQ602215S, RSQ602282S, RSQ602328S, RSQ602361S, RSQ601844S, RSQ601996S, RSQ601998S, RSQ602212S, RSQ602283S, RSQ601829S, RSQ601830S, RSQ601832S, RSQ601833S, RSQ601836S, RSQ601991S, RSQ601992S, RSQ601993S, RSQ601994S, RSQ601995S, RSQ602195S, RSQ602196S, RSQ602197S, RSQ602198S, RSQ602199S, RSQ602200S, RSQ602201S, RSQ602202S, RSQ602203S, RSQ602204S, RSQ602470S, RSQ602488S, RSQ602489S, RSQ602491S, RSQ602492S, RSQ600706S, RSQ601019S, RSQ601233S, RSQ601298S, RSQ601306S, RSQ601581S, RSQ601849S, RSQ602284S, RSQ602358S, RSQ602520S, RSQ602577S, RSQ601317S, RSQ601626S, RSQ601633S, RSQ601640S, RSQ601642S, RSQ601645S, RSQ601654S, RSQ601655S, RSQ601656S, RSQ601657S, RSQ601460S, RSQ601461S, RSQ601462S, RSQ601463S, RSQ601464S, RSQ601465S, RSQ601466S, RSQ601467S, RSQ601468S, RSQ601469S, RSQ601470S, RSQ601471S, RSQ601472S, RSQ601475S, RSQ601477S, RSQ601478S, RSQ601494S, RSQ601495S, RSQ601496S, RSQ601497S, RSQ601498S, RSQ601499S, RSQ601500S, RSQ601502S, RSQ601504S, RSQ601505S, RSQ601507S, RSQ601509S, RSQ601510S, RSQ601513S, RSQ600230S, RSQ600587S, RSQ600011S, RSQ600013S, RSQ600838S, RSQ602521S, RSQ600597S, RSQ600711S, RSQ600886S, RSQ601241S, RSQ601308S, RSQ602154S, RSQ602347S, RSQ600260S, RSQ600206S, RSQ600215S, RSQ600217S, RSQ600434S, RSQ600730S, RSQ600872S, RSQ600365S, RSQ600366S, RSQ600007S, RSQ600009S, RSQ600014S, RSQ600015S, RSQ600383S, RSQ600387S, RSQ600488S, RSQ600512S, RSQ600518S, RSQ600579S, RSQ600853S, RSQ600880S, RSQ600882S, RSQ600888S, RSQ600889S, RSQ600006S, RSQ600018S, RSQ600475S, RSQ600487S, RSQ600578S, RSQ600590S, RSQ600857S, RSQ601089S, RSQ601217S, RSQ601250S, RSQ601842S, RSQ600551S, RSQ600864S, RSQ600865S, RSQ601126S, RSQ601245S, RSQ601400S, RSQ601587S, RSQ601665S, RSQ601922S, RSQ602342S, RSQ602363S, RSQ602385S, RSQ602302S, RSQ600492S, RSQ600008S, RSQ600023S, RSQ600519S, RSQ600588S, RSQ600714S, RSQ600893S, RSQ600696S, RSQ600697S, RSQ601309S, RSQ600856S, RSQ600862S, RSQ602380S, RSQ601230S, RSQ602354S, RSQ602369S, RSQ602507S, RSQ602522S, RSQ601098S, RSQ600468S, RSQ600189S, RSQ600600S, RSQ601248S, RSQ601314S, RSQ600598S, RSQ600892S, RSQ601257S, RSQ601385S, RSQ601791S, RSQ601916S, RSQ602348S, RSQ602384S, RSQ600473S, RSQ601409S, RSQ601748S, RSQ601749S, RSQ601751S, RSQ600044S, RSQ600045S, RSQ600193S, RSQ600250S, RSQ600340S, RSQ600441S, RSQ600514S, RSQ600516S, RSQ600517S, RSQ600563S, RSQ600564S, RSQ600568S, RSQ600599S, RSQ600603S, RSQ600615S, RSQ601025S, RSQ601074S, RSQ601076S, RSQ601095S, RSQ601399S, RSQ601595S, RSQ601904S, RSQ602209S, RSQ602531S, RSQ600386S, RSQ600393S, RSQ600396S, RSQ600397S, RSQ600398S, RSQ600399S, RSQ600400S, RSQ600401S, RSQ600402S, RSQ600403S, RSQ600404S, RSQ600405S, RSQ600406S, RSQ600407S, RSQ600412S, RSQ600413S, RSQ600414S, RSQ600440S, RSQ600442S, RSQ600469S, RSQ600486S, RSQ600498S, RSQ600503S, RSQ600504S, RSQ600515S, RSQ600520S, RSQ600521S, RSQ600527S, RSQ600555S, RSQ600583S, RSQ601020S, RSQ601059S, RSQ601106S, RSQ601117S, RSQ601121S, RSQ601128S, RSQ601133S, RSQ601144S, RSQ601145S, RSQ601240S, RSQ601249S, RSQ601319S, RSQ601320S, RSQ601352S, RSQ601355S, RSQ601358S, RSQ601360S, RSQ601365S, RSQ601388S, RSQ601389S, RSQ601517S, RSQ601518S, RSQ601520S, RSQ601521S, RSQ601522S, RSQ601531S, RSQ601532S, RSQ601544S, RSQ601560S, RSQ601772S, RSQ601773S, RSQ601779S, RSQ601780S, RSQ601781S, RSQ601782S, RSQ601815S, RSQ601825S, RSQ601831S, RSQ601834S, RSQ601865S, RSQ601889S, RSQ601915S, RSQ601940S, RSQ601947S, RSQ601948S, RSQ601951S, RSQ601953S, RSQ601958S, RSQ602061S, RSQ602063S, RSQ602066S, RSQ602067S, RSQ602071S, RSQ602097S, RSQ602100S, RSQ602103S, RSQ602105S, RSQ602110S, RSQ602112S, RSQ602125S, RSQ602131S, RSQ602133S, RSQ602147S, RSQ602155S, RSQ602157S, RSQ602158S, RSQ602167S, RSQ602173S, RSQ602187S, RSQ602264S, RSQ602375S, RSQ602376S, RSQ600408S, RSQ600409S, RSQ600448S, RSQ600522S, RSQ600616S, RSQ601047S, RSQ601115S, RSQ601274S, RSQ601304S, RSQ601361S, RSQ601540S, RSQ601778S, RSQ602069S, RSQ602124S, RSQ602186S, RSQ600684S, RSQ600909S, RSQ600999S, RSQ601271S, RSQ601332S, RSQ601368S, RSQ602101S, RSQ600672S, RSQ601285S, RSQ601333S, RSQ601367S, RSQ602188S, RSQ602127S, RSQ601525S, RSQ601526S, RSQ601527S, RSQ601528S, RSQ601530S, RSQ600625S, RSQ600629S, RSQ600630S, RSQ600632S, RSQ600636S, RSQ600638S, RSQ601140S, RSQ601253S, RSQ601263S, RSQ601275S, RSQ601276S, RSQ601327S, RSQ601330S, RSQ601362S, RSQ601364S, RSQ601371S, RSQ601402S, RSQ601432S, RSQ601501S, RSQ601537S, RSQ601543S, RSQ601545S, RSQ601789S, RSQ601799S, RSQ601803S, RSQ601804S, RSQ601817S, RSQ601888S, RSQ601891S, RSQ601917S, RSQ601929S, RSQ601944S, RSQ601955S, RSQ602102S, RSQ602107S, RSQ602141S, RSQ602153S, RSQ602174S, RSQ602178S, RSQ602579S, RSQ600617S, RSQ600619S, RSQ600623S, RSQ601054S, RSQ601109S, RSQ601130S, RSQ601262S, RSQ601268S, RSQ601278S, RSQ601279S, RSQ601328S, RSQ601331S, RSQ601370S, RSQ601397S, RSQ601523S, RSQ601524S, RSQ601538S, RSQ601539S, RSQ601776S, RSQ601801S, RSQ601808S, RSQ602068S, RSQ602129S, RSQ602175S, RSQ602177S, RSQ602182S, RSQ600450S, RSQ600451S, RSQ600452S, RSQ600455S, RSQ600456S, RSQ600463S, RSQ600474S, RSQ600642S, RSQ600644S, RSQ600680S, RSQ600689S, RSQ601003S, RSQ601005S, RSQ601027S, RSQ601035S, RSQ601050S, RSQ601051S, RSQ601060S, RSQ601102S, RSQ601118S, RSQ601125S, RSQ601136S, RSQ601142S, RSQ601254S, RSQ601255S, RSQ601256S, RSQ601284S, RSQ601294S, RSQ601295S, RSQ601303S, RSQ601325S, RSQ601329S, RSQ601366S, RSQ601431S, RSQ601548S, RSQ601551S, RSQ601552S, RSQ601558S, RSQ601798S, RSQ601807S, RSQ601809S, RSQ601857S, RSQ601875S, RSQ601878S, RSQ602122S, RSQ602132S, RSQ602136S, RSQ602165S, RSQ602467S, RSQ602508S, RSQ602515S, RSQ600370S, RSQ600371S, RSQ600372S, RSQ600373S, RSQ600379S, RSQ600411S, RSQ600447S, RSQ601030S, RSQ601124S, RSQ601139S, RSQ601141S, RSQ601234S, RSQ601354S, RSQ601398S, RSQ601785S, RSQ601819S, RSQ602104S, RSQ600612S, RSQ600613S, RSQ600614S, RSQ601131S, RSQ601251S, RSQ601292S, RSQ601353S, RSQ601356S, RSQ601516S, RSQ601561S, RSQ601792S, RSQ601802S, RSQ601828S, RSQ601838S, RSQ602163S, RSQ602181S, RSQ601072S, RSQ602568S, RSQ601739S, RSQ600026S, RSQ600392S, RSQ600454S, RSQ600457S, RSQ600526S, RSQ600634S, RSQ600875S, RSQ600876S, RSQ600956S, RSQ600961S, RSQ600963S, RSQ600967S, RSQ601012S, RSQ601013S, RSQ601572S, RSQ601659S, RSQ601707S, RSQ601736S, RSQ601867S, RSQ601913S, RSQ601990S, RSQ602319S, RSQ602344S, RSQ602505S, RSQ602553S, RSQ601099S, RSQ601712S, RSQ601726S, RSQ600196S, RSQ600500S, RSQ600502S, RSQ600539S, RSQ600541S, RSQ600542S, RSQ600544S, RSQ600547S, RSQ600699S, RSQ600724S, RSQ601075S, RSQ601119S, RSQ601200S, RSQ601204S, RSQ601417S, RSQ601418S, RSQ601484S, RSQ601487S, RSQ600187S, RSQ600203S, RSQ600234S, RSQ600388S, RSQ600390S, RSQ600810S, RSQ601488S, RSQ600041S, RSQ601173S, RSQ601184S, RSQ601489S, RSQ601491S, RSQ601715S, RSQ601716S, RSQ601717S, RSQ601718S, RSQ601721S, RSQ601722S, RSQ601759S, RSQ601761S, RSQ601762S, RSQ601926S, RSQ601973S, RSQ601975S, RSQ601977S, RSQ601980S, RSQ601981S, RSQ602252S, RSQ602262S, RSQ600509S, RSQ601481S, RSQ600510S, RSQ602260S, RSQ600391S, RSQ601084S, RSQ600548S, RSQ600508S, RSQ601180S, RSQ600947S, RSQ600499S, RSQ601444S, RSQ601861S, RSQ600495S, RSQ600874S, RSQ600543S, RSQ601187S, RSQ601195S, RSQ600497S, RSQ601713S, RSQ601714S, RSQ600364S, RSQ600369S, RSQ600376S, RSQ600377S, RSQ600483S, RSQ600540S, RSQ600818S, RSQ600870S, RSQ601100S, RSQ601676S, RSQ601681S, RSQ601688S, RSQ601723S, RSQ601724S, RSQ601728S, RSQ600536S, RSQ600537S, RSQ600538S, RSQ600545S, RSQ600546S, RSQ601039S, RSQ601482S, RSQ601483S, RSQ601485S, RSQ601486S, RSQ600493S, RSQ600496S, RSQ600505S, RSQ600506S, RSQ600507S, RSQ600549S, RSQ601070S, RSQ601120S, RSQ601197S, RSQ601206S, RSQ601207S, RSQ601209S, RSQ601480S, RSQ601685S, RSQ601711S, RSQ601719S, RSQ601720S, RSQ601928S, RSQ601931S, RSQ601932S, RSQ601984S, RSQ601985S, RSQ602249S, RSQ600034S, RSQ600491S, RSQ600513S, RSQ600715S, RSQ600589S, RSQ600887S, RSQ601065S, RSQ601083S, RSQ601242S, RSQ600367S, RSQ600368S, RSQ600378S, RSQ600584S, RSQ600682S, RSQ600692S, RSQ600898S, RSQ601111S, RSQ601113S, RSQ601556S, RSQ601765S, RSQ601774S, RSQ601775S, RSQ601805S, RSQ601814S, RSQ601882S, RSQ600038S, RSQ600039S, RSQ600524S, RSQ601816S, RSQ601571S, RSQ600010S, RSQ600490S, RSQ600511S, RSQ601033S, RSQ600433S, RSQ600470S, RSQ600529S, RSQ601750S, RSQ601756S, RSQ601810S, RSQ601820S, RSQ602227S, RSQ600040S, RSQ600198S, RSQ601064S, RSQ600380S, RSQ600384S, RSQ600385S, RSQ600410S, RSQ600482S, RSQ600501S, RSQ600681S, RSQ601269S, RSQ601305S, RSQ601311S, RSQ601324S, RSQ601336S, RSQ601372S, RSQ601429S, RSQ601511S, RSQ601542S, RSQ601763S, RSQ601764S, RSQ601771S, RSQ601784S, RSQ601786S, RSQ601787S, RSQ601788S, RSQ601813S, RSQ601868S, RSQ601873S, RSQ601893S, RSQ601905S, RSQ601969S, RSQ601972S, RSQ602059S, RSQ602113S, RSQ602152S, RSQ602161S, RSQ602162S, RSQ601405S, RSQ601408S, RSQ600020S, RSQ600731S, RSQ600879S, RSQ601252S, RSQ601843S, RSQ601310S, RSQ601449S, RSQ600030S, RSQ600471S, RSQ600572S, RSQ600624S, RSQ600651S, RSQ600840S, RSQ600869S, RSQ601038S, RSQ601045S, RSQ601069S, RSQ601146S, RSQ601236S, RSQ601237S, RSQ601273S, RSQ601412S, RSQ601426S, RSQ601436S, RSQ601570S, RSQ601624S, RSQ601746S, RSQ601899S, RSQ602317S, RSQ600646S, RSQ600649S, RSQ600654S, RSQ600659S, RSQ600661S, RSQ601034S, RSQ601037S, RSQ601143S, RSQ601149S, RSQ601280S, RSQ601283S, RSQ601291S, RSQ601316S, RSQ601323S, RSQ601326S, RSQ601334S, RSQ601335S, RSQ601357S, RSQ601407S, RSQ601423S, RSQ601428S, RSQ601493S, RSQ601508S, RSQ601770S, RSQ601790S, RSQ601822S, RSQ601835S, RSQ601894S, RSQ601979S, RSQ602108S, RSQ602120S, RSQ602121S, RSQ602148S, RSQ602189S, RSQ602360S, RSQ602496S, RSQ602523S, RSQ600574S, RSQ600037S, RSQ600394S, RSQ600484S, RSQ600565S, RSQ600580S, RSQ600596S, RSQ600883S, RSQ600884S, RSQ601022S, RSQ601066S, RSQ601092S, RSQ601215S, RSQ601247S, RSQ601259S, RSQ601735S, RSQ602300S, RSQ601213S, RSQ600458S, RSQ600461S, RSQ600476S, RSQ600477S, RSQ600479S, RSQ600480S, RSQ600523S, RSQ600525S, RSQ600534S, RSQ600609S, RSQ600673S, RSQ600901S, RSQ600902S, RSQ601031S, RSQ601049S, RSQ601056S, RSQ601104S, RSQ601105S, RSQ601112S, RSQ601116S, RSQ601123S, RSQ601129S, RSQ601243S, RSQ601261S, RSQ601281S, RSQ601286S, RSQ601301S, RSQ601302S, RSQ601318S, RSQ601321S, RSQ601322S, RSQ601359S, RSQ601363S, RSQ601506S, RSQ601515S, RSQ601557S, RSQ601794S, RSQ601796S, RSQ601797S, RSQ601800S, RSQ601818S, RSQ601824S, RSQ601890S, RSQ601906S, RSQ601920S, RSQ601950S, RSQ601954S, RSQ601967S, RSQ601978S, RSQ602055S, RSQ602064S, RSQ602077S, RSQ602083S, RSQ602085S, RSQ602098S, RSQ602099S, RSQ602109S, RSQ602111S, RSQ602115S, RSQ602144S, RSQ602146S, RSQ602156S, RSQ602164S, RSQ602166S, RSQ602183S, RSQ602378S, RSQ602529S, RSQ602569S, RSQ600264S, RSQ600019S, RSQ601514S, RSQ601519S, RSQ601529S, RSQ601533S, RSQ601534S, RSQ601535S, RSQ601541S, RSQ601821S, RSQ602387S, RSQ602388S, RSQ602389S, RSQ602390S, RSQ602391S, RSQ602392S, RSQ602394S, RSQ600594S, RSQ600861S, RSQ600558S, RSQ601755S, RSQ601044S, RSQ601512S, RSQ600868S, RSQ601576S, RSQ601623S, RSQ601692S, RSQ601705S, RSQ601706S, RSQ601727S, RSQ601743S, RSQ601744S, RSQ601760S, RSQ601769S, RSQ601783S, RSQ601845S, RSQ601852S, RSQ601863S, RSQ601943S, RSQ601945S, RSQ601946S, RSQ601949S, RSQ601956S, RSQ601959S, RSQ602030S, RSQ602065S, RSQ602076S, RSQ602114S, RSQ602116S, RSQ602117S, RSQ602119S, RSQ602123S, RSQ602126S, RSQ602137S, RSQ602140S, RSQ602145S, RSQ602149S, RSQ602159S, RSQ602160S, RSQ602168S, RSQ602170S, RSQ602172S, RSQ602185S, RSQ602205S, RSQ602228S, RSQ602230S, RSQ602237S, RSQ602238S, RSQ602239S, RSQ602248S, RSQ602250S, RSQ602263S, RSQ602265S, RSQ602286S, RSQ602288S, RSQ602291S, RSQ602292S, RSQ602293S, RSQ602316S, RSQ602322S, RSQ602323S, RSQ602329S, RSQ602330S, RSQ602332S, RSQ602341S, RSQ602362S, RSQ602371S, RSQ602377S, RSQ602469S, RSQ602471S, RSQ602474S, RSQ602481S, RSQ602495S, RSQ602497S, RSQ602502S, RSQ602504S, RSQ602510S, RSQ602511S, RSQ602513S, RSQ602519S, RSQ602528S, RSQ602530S, RSQ602533S, RSQ602541S, RSQ602542S, RSQ602544S, RSQ602546S, RSQ602551S, RSQ602554S, RSQ602558S, RSQ602567S, RSQ602571S, RSQ602573S, RSQ602574S, RSQ602578S, RSQ602585S, RSQ602590S, RSQ602594S, RSQ602596S, RSQ602597S, RSQ602598S, RSQ602602S, RSQ602605S, RSQ602606S, RSQ602607S, RSQ602608S, RSQ602609S, RSQ602610S, RSQ602611S, RSQ602612S, RSQ602613S, RSQ602614S, RSQ602615S, RSQ602616S, RSQ602619S, RSQ602620S, RSQ602621S, RSQ602622S, RSQ602643S, RSQ602644S, RSQ602645S, RSQ602646S, RSQ602647S, RSQ602648S, RSQ602649S, RSQ602650S, RSQ602651S, RSQ602652S, RSQ602653S, RSQ602654S, RSQ602655S, RSQ602656S, RSQ602658S, RSQ602657S, RSQ601238S, RSQ601710S, RSQ601777S, RSQ602138S, RSQ602179S, RSQ602232S, RSQ602251S, RSQ602266S, RSQ602512S, RSQ602552S, RSQ602559S, RSQ602570S",
      "more_code_info": ""
    }
  ]
}