  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65670",
      "recalling_firm": "Synthes USA HQ, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including PA, CT, DE, and NY.",
      "recall_number": "Z-1726-2013",
      "product_description": "Synthes CMF Distraction System    The product is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus, where gradual bone distraction is required. It is intended for single use only.",
      "product_quantity": "10",
      "reason_for_recall": "The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF)  Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots in which the slot in the barrel of the BC distractor was rotated such that it was not possible to attach the B-type and C-type foot plates.",
      "recall_initiation_date": "20130612",
      "center_classification_date": "20130712",
      "termination_date": "20150921",
      "report_date": "20130724",
      "code_info": "Part # 04.315.067 with lot #s 6184270 and/or 22612-04."
    }
  ]
}