  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98598",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).",
      "recall_number": "Z-1721-2026",
      "product_description": "Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:    1. HEAD AND NECK CDS-LF, Medline SKU # CDS780138K;   2. HEART CATH PACK-LF, Medline SKU # DYNJ36478B;   3. HEART CATH PACK-LF, Medline SKU # DYNJ36478B;   4. HEART CATH PACK-LF, Medline SKU # DYNJ36478B;   5. THYROIDECTOMY, Medline SKU # DYNJ910458;   6. LH 3 PORT W/O WASTEBAG, Medline SKU # VASC1081C;   7. LHK, Medline SKU # VASC1510;   8. LHK, Medline SKU # VASC1510.",
      "product_quantity": "594 kits",
      "reason_for_recall": "Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.",
      "recall_initiation_date": "20260227",
      "center_classification_date": "20260407",
      "report_date": "20260415",
      "code_info": "Medline SKU # CDS780138K, UDI/DI each 10198459205552, UDI/DI case 40198459205553, Lot Number: 25EBR802;  Medline SKU # DYNJ36478B, UDI/DI each 10889942814619, UDI/DI case 40889942814610, Lot Number: 25AMH487;  Medline SKU # DYNJ36478B, UDI/DI each 10889942814619, UDI/DI case 40889942814610, Lot Number: 24JMC622;  Medline SKU # DYNJ36478B, UDI/DI each 10889942814619, UDI/DI case 40889942814610, Lot Number: 24EMD869;  Medline SKU # DYNJ910458, UDI/DI each 10195327692469, UDI/DI case 40195327692460, Lot Number: 24LLA039;  Medline SKU # VASC1081C, UDI/DI each 10195327054519, UDI/DI case 40195327054510, Lot Number: 25HBU295;  Medline SKU # VASC1510, UDI/DI each 10195327686888, UDI/DI case 40195327686889, Lot Number: 24GBO586;  Medline SKU # VASC1510, UDI/DI each 10195327686888, UDI/DI case 40195327686889, Lot Number: 24EBG342.",
      "more_code_info": ""
    }
  ]
}