{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94276",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.",
      "recall_number": "Z-1719-2024",
      "product_description": "ZPLP (Zimmer Periarticular Locking Plate System)  Distal Lateral Fibular Plate, Left, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures  Item Number: 47-2357-018-12",
      "product_quantity": "",
      "reason_for_recall": "Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention",
      "recall_initiation_date": "20240320",
      "center_classification_date": "20240430",
      "report_date": "20240508",
      "code_info": "UDI:  (01)00889024088603(17)290512(10)64381071 (01)00889024088603(17)310324(10)65038410 (01)00889024088603(17)310324(10)65038411 (01)00889024088603(17)310324(10)65038412 (01)00889024088603(17)310821(10)65193244 (01)00889024088603(17)311018(10)65233124 (01)00889024088603(17)311017(10)65233125  Lot Numbers:  64381071 65038410 65038411 65038412 65193244 65233124 65233125",
      "more_code_info": ""
    }
  ]
}