  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65459",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.",
      "recall_number": "Z-1718-2013",
      "product_description": "REF 154231Oxford Fixed Bearing Partial Knee  Cemented Lateral Tibia  Size E6 ARCOM UHMWPE / CO-CR-MO Alloy  Only for use in Lateral compartment  Sterile R    The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.",
      "product_quantity": "28",
      "reason_for_recall": "Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.",
      "recall_initiation_date": "20130607",
      "center_classification_date": "20130711",
      "termination_date": "20140717",
      "report_date": "20130717",
      "code_info": "M321600, M539430, M444910, M657070,M722440, M136980, M444910"
    }
  ]
}