{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98530",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.",
      "recall_number": "Z-1710-2026",
      "product_description": "Medline kits containing Olympus biopsy valves:    1. BRONCH PROCEDURE KIT, DYKE2091    2. BRONCHSCOPY, DYNJ900898G DYNJ901922I      3. BRONCHSCOPY PACK, DYNJ38313B    4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I    5. KIT, WR THORACIC, DYNJ911252    6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B    7. LEX THORACIC, DYNJ902016K DYNJ902016L    8. ORGAN PROCUREMENT, DYNJ908686    9. PULMONARY PACK, DYKE1859    10. THORACIC, DYNJ901666L    11. THORACIC ROBOTICS, DYNJ908777A    12. THORACOTOMY/THORACOSCOPY, DYNJ900482N DYNJ900482O    13. XPS, DYNJ907605",
      "product_quantity": "14,379 kits total",
      "reason_for_recall": "Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.",
      "recall_initiation_date": "20260216",
      "center_classification_date": "20260401",
      "report_date": "20260408",
      "code_info": "1. DYKE2091 UDI-DI 10198459606397 (EA) 40198459606398 (CS) LOT 25LBR616  2. DYNJ900898G UDI-DI 10195327479473 (EA) 40195327479474 (CS) LOTS 23JBI824 23KBJ535 24ABR789 24ABV917 24BBU610 24DMK530 24FMF039 24GME122 24JMB670 24KMH657 25AMA403 25BME938  DYNJ901922I UDI-DI 10198459237270 (EA) 40198459237271 (CS) LOTS 25BBL628 25FBM655 25HBW072 25KBO230  3. DYNJ38313B UDI-DI 10884389857700 (EA) 40884389857701 (CS) LOTS 24BBR676 24CBF032 24FBD924 24JBE567 25EBA277 25FBO580 25IBI418  4. DYNJ32750G UDI-DI 10889942212491 (EA) 40889942212492 (CS) LOTS 23KME954 23LMD492 24BMF722 24FMH267 24HMC509  DYNJ32750I UDI-DI 10198459245695 (EA) 40198459245696 (CS) LOTS 25BMA765 25DMC474 25EMI991 25IME718 26AMJ190  5. DYNJ911252 UDI-DI 10198459305030 (EA) 40198459305031 (CS) LOTS 25EBU007 25LBC286 25LBI365 26ABC803 26ABU372  6. DYNJ58146B UDI-DI 10198459478253 (EA) 40198459478254 (CS) LOTS 25HMG837 25JMJ636 26AMJ160  7. DYNJ902016K UDI-DI 10195327391706 (EA) 40195327391707 (CS) LOTS 23JBL313 23KBX206 24ABG361 24ABS317 24CBU624 24DBT777 24EBP517 24IBB815 24LBH029 24LBK316 25BBD830 25BBK993 25DBQ241 25DBR926  DYNJ902016L UDI-DI 10198459452703 (EA) 40198459452704 (CS) LOTS 25HBV912 25IBT298 25KBT137  8. DYNJ908686 UDI-DI 10195327145330 (EA) 40195327145331 (CS) LOT 23LBG479  9. DYKE1859 UDI-DI 10195327171469 (EA) 40195327171460 (CS) LOTS 23LBU165 24ABT730 24CBL345 24FBK106 24HBJ512 25DBS024 25EBL346  10. DYNJ901666L UDI-DI 10193489986815 (EA) 40193489986816 (CS) LOTS 24DDC241 24EDB399  11. DYNJ908777A UDI-DI 10195327415174 (EA) 40195327415175 (CS) LOTS 23JBV709 23KBC057 23LBP026 23LBP071 23LBP286 24BBH036 24DBD851 24FBC615 24GBH920 24HBS270 24KBL822 25ABE547 25BBP040 25BBR359 25CBA577  12. DYNJ900482N UDI-DI 10195327509941 (EA) 40195327509942 (CS) LOTS 23KDB600 24ADA178  DYNJ900482O UDI-DI 10195327596569 (EA) 40195327596560 (CS) LOTS 24CDA456 24EDC051 24GDB830 24IDB049 24JDB140 24KDA731 25ADB869 25BDA267 25CMD906  13. DYNJ907605 UDI-DI 10193489468052 (EA) 40193489468053 (CS) LOTS 23JBA743 23JBI519 23LBQ647 24ABX014 24BBQ594",
      "more_code_info": ""
    }
  ]
}