  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94276",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.",
      "recall_number": "Z-1709-2024",
      "product_description": "ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures  Item Number: 47-2357-017-04",
      "product_quantity": "",
      "reason_for_recall": "Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention",
      "recall_initiation_date": "20240320",
      "center_classification_date": "20240430",
      "report_date": "20240508",
      "code_info": "UDI:  (01)00889024088498(17)321205(10)65620069 (01)00889024088498(17)321205(10)65635922 (01)00889024088498(17)321205(10)65635923 (01)00889024088498(17)330111(10)65674071 (01)00889024088498(17)330115(10)65674072 (01)00889024088498(17)330428(10)66010704 (01)00889024088498(17)330730(10)66166096 (01)00889024088498(17)330822(10)66216633  Lot Numbers: 65620069 65635922 65635923 65674071 65674072 66010704 66166096 66216633",
      "more_code_info": ""
    }
  ]
}