{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98461",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.",
      "recall_number": "Z-1703-2026",
      "product_description": "2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.",
      "product_quantity": "33 units",
      "reason_for_recall": "A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.",
      "recall_initiation_date": "20260225",
      "center_classification_date": "20260330",
      "report_date": "20260408",
      "code_info": "Part Number: 03.424.126-US. GTIN: 10886982335951. Lot Number: 68982P8.",
      "more_code_info": ""
    }
  ]
}