{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94375",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.",
      "recall_number": "Z-1700-2024",
      "product_description": "POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic.    Model/Catalog Number:  PS-0537CJDA",
      "product_quantity": "850 units: (400 units US);(450 units OUS)",
      "reason_for_recall": "Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.",
      "recall_initiation_date": "20240307",
      "center_classification_date": "20240430",
      "report_date": "20240508",
      "code_info": "UDI-DI: 00821925044555 Lot numbers:  CA182543, CA182544, CA182546, CA182548, CA191230",
      "more_code_info": ""
    }
  ]
}