  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mettawa",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96664",
      "recalling_firm": "Vyaire Medical",
      "address_1": "26125 N Riverwoods Blvd",
      "address_2": "",
      "postal_code": "60045-3420",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.",
      "recall_number": "Z-1698-2025",
      "product_description": "Brand Name: AirLife  Product Name: AirLife Infant Heated Wire Circuit  Model/Catalog Number: AH265  Software Version: N/A  Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber  Component: No. a conduit for respiratory gas between the patient and a ventilator.",
      "product_quantity": "3053",
      "reason_for_recall": "Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.",
      "recall_initiation_date": "20250410",
      "center_classification_date": "20250509",
      "report_date": "20250521",
      "code_info": "Lot Code: Lot/Serial Number(s): 0004247924   0004247925   0004252940   0004252941   0004256292   0004256293   0004256294   0004256295  0004272325 0004272868  0004278978  0004279156 0004280036 0004285304  0004288036  0004289514   UDI-DI:  AirLife Label Each: 10889483595909 Case: 30889483595903  Vyaire Label: Each: 10190752145160 Case: 50190752145168",
      "more_code_info": ""
    }
  ]
}