  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Norwood",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94330",
      "recalling_firm": "Advanced Instruments, LLC",
      "address_1": "2 Technology Way",
      "address_2": "",
      "postal_code": "02062-2633",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic,  France, Germany, Iceland, Netherlands, South Korea, Spain, Switzerland, Taiwan.",
      "recall_number": "Z-1696-2024",
      "product_description": "OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine.  Model/Catalog Number: OsmoPRO MAX   Software Version: 1.5.0.0",
      "product_quantity": "58 units",
      "reason_for_recall": "System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.",
      "recall_initiation_date": "20240311",
      "center_classification_date": "20240429",
      "report_date": "20240508",
      "code_info": "UDI-DI: 00816068021150;  Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A.  All serial numbers that end in  A",
      "more_code_info": ""
    }
  ]
}