  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sligo",
      "state": "",
      "country": "Ireland",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92271",
      "recalling_firm": "Abbott Ireland Limited",
      "address_1": "Finisklin Business & Technology Park",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, CO, FL, GA, IL, KY, LA, MD, MI, MN, NJ, NY, OH, RI, SC, SD, TN, TX, WA.  No international distribution of affected lot.",
      "recall_number": "Z-1696-2023",
      "product_description": "Alinity i Anti-TPO Reagent Kit, List Number 09P3521",
      "product_quantity": "110 kits",
      "reason_for_recall": "Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU).",
      "recall_initiation_date": "20230424",
      "center_classification_date": "20230608",
      "report_date": "20230614",
      "code_info": "UDI Number: (01)00380740153892(17)240720(10)46009FN00; List Number: 09P3521; Lot Number: 46009FN00; Expiration Date: 07/20/2024",
      "more_code_info": ""
    }
  ]
}