{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98422",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.",
      "recall_number": "Z-1695-2026",
      "product_description": "Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System  Product Name: SOLTIVE Pro Laser System TFL-SLS  Model/Catalog Number: TFL-SLS containing TFL-CSLU  Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis.  Component: No",
      "product_quantity": "32 units",
      "reason_for_recall": "Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.",
      "recall_initiation_date": "20260227",
      "center_classification_date": "20260330",
      "report_date": "20260408",
      "code_info": "Laser System:  TFL-SLS; UDI:  00821925044135;         Component Part: TFL-CSLU; UDI:  00821925044593;   Serial #: MDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164, MDUF230166, MDUF230167, MDUF230234, MDUF230283, MDUF230284, MDUF230331, MDUF230332, MDUF230361, MDUF230362, MDUF230363, MDUF230364, MDUF230365, MDUF230366, MDUF230411, MDUF230412, MDUF230413;",
      "more_code_info": ""
    }
  ]
}