  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65590",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including UT, CA, AZ, CO, FL, HI. MT, NY,TX and WA and one customer in Russia.",
      "recall_number": "Z-1686-2013",
      "product_description": "da Vinci Si Surgeon Side Cart, model number SS3000;    Manufactured by Intuitive Surgical, Sunnyvale, CA 94086.    Intuitive Surgical endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic instruments. It is intended for adult and pediatric use.",
      "product_quantity": "30 total, all devices",
      "reason_for_recall": "Factory testing on da Vinci Si Surgical Systems may not be in compliance with UL Standards as one of the testing devices was found to be working incorrectly.",
      "recall_initiation_date": "20130608",
      "center_classification_date": "20130711",
      "termination_date": "20140122",
      "report_date": "20130717",
      "code_info": "Part number 380610-15:  Serial numbers:   323936  323937  325082  325143  325175  325178  325355  325290  325352  325669  325672  325692  325444  325446  325447  325583  325585  325587  325634  325671  325684  325720  325724  325719  325888  325465  325549  325266  325268  323875  324092  324098  324100  324190  324258  324260  324265  324370  324425  324426  324528  324585  324586  324638  324649  324847  324848  324849  325076"
    }
  ]
}