      "country": "United States",
      "city": "Sunnyvale",
      "address_1": "1310 Moffett Park Dr",
      "reason_for_recall": "Software anomaly on OptiMedica CATALYS System version 3.00.05 that may result in incorrect parameters provided for cataract incision surgeon templates and both eyes have the same parameters.",
      "address_2": "",
      "product_quantity": "171 distributed Worldwide",
      "code_info": "Software version 3.00.05:  Serial numbers - US:   44066413  44063113  44065413  44066213  44102714  44060012  44064813  44061012  44061812  44101614  44068314  44056612  44063713  44102614  44068914  44073514  44074814  44061612  44073714  44056312  44061412  44066613  44069414  44069814  44070914  44072714  44073014  44073114  44074514  44074614  44100714  44068014  44060612  44062813  44066914  44056012  44062213  44064513  44101014  44100814  44060312  44060412  44063513  44056712  44063313  44063913  44073314  44065613  44058312  44065013  44072114  44072414  44071914  44062613  44065213  44059912  44057312  44059712  44058212  44072814  44060712  44060112  44057812  44058012  44059612  44069114    Serial numbers - OUS:   44055211  44055411  44055711  44056112  44056212  44056412  44056512  44056912  44057112  44057212  44057412  44057512  44057712  44057912  44058512  44058612  44058912  44059012  44059112  44059412  44059512  44061212  44061312  44061512  44061712  44061913  44062113  44062313  44063213  44064013  44064213  44065513  44066313  44066813  44067214  44067414  44067514  44067714  44068114  44068414  44068614  44069214  44069514  44069914  44070014  44070114  44070614  44070714  44071014  44071214  44071314  44071414  44071514  44071714  44071814  44072214  44072314  44072614  44073414  44074014  44074214  44074414  44074714  44074914  44100114  44100614  44101714  44101814  44101914  44102014.",
      "center_classification_date": "20150528",
      "distribution_pattern": "Worldwide Distribution - US Nationwide, Germany, Switzerland, Australia, Spain, Japan, Great Britain, France, Netherlands, Austria, Singapore, South Korea, India, Brazil, Italy, Belgium, Egypt, Colombia, Hong Kong, and Canada.",
      "state": "CA",
      "product_description": "Catalys Precision Laser System-Catalys - U - US product; Catalys-1 - International product; software version 3.00.05    The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing  cataract surgery for removal of the crystalline lens.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Optimedica Corporation",
      "recall_number": "Z-1683-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70672",
      "termination_date": "20151208",
      "more_code_info": "",
      "recall_initiation_date": "20150220",
      "postal_code": "94089-1133",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}