  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bothell",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85314",
      "recalling_firm": "Solta Medical Inc",
      "address_1": "11720 N Creek Pkwy N Ste 100",
      "address_2": "N/A",
      "postal_code": "98011-8244",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US - AK, AR, AZ, CA, FL, GA, IL, NY, PA, TX. UT, and WA    OUS -Japan, Korea, Malaysia, Singapore, Taiwan, and Thailand.",
      "recall_number": "Z-1681-2020",
      "product_description": "Thermage Face Tip 3.0, REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF TTNS3.00C3-600/UDI (01)000850608002384; REF TTNS3.00C3-900; UDI (01)000850608002391  Used in dermatologic and general surgical procedures for electrocoagulation and hemostasis.",
      "product_quantity": "4472 units",
      "reason_for_recall": "Complaints of sparking of the tips of the electrosurgical device used in RF surgical procedures. The use of the affected treatment tips could contribute to potential skin burns.",
      "recall_initiation_date": "20191230",
      "center_classification_date": "20200407",
      "termination_date": "20240429",
      "report_date": "20200415",
      "code_info": "Model Number /Lot Numbers:  TTNS3.00C3-400: DOM 120519 011,  DOM 120419 014, DOM 120219 008, DOM 112619 006, DOM 112019 010, DOM 111519 004, DOM 100819 005, DOM 100119 005, DOM 082319 003, DOM 081919 007, DOM 081419 004;     TTNS3.00C3-600: DOM 120419 018, DOM 112019 009, DOM 110119 010, DOM 092319 005, DOM 082219 015, DOM 082119 006;     TTNS3.00C3-900: DOM 111819 005, DOM 092319 006  , DOM 081419 005    Lot number/Number of Units Affected  120219-008/180, 112619-006/225, 112019-010/225, 112019-009/420, 111519-004/220, 110119-010/398, 100819-005/405, 092319-006/23, 092319-005/342,  082319-003/525, 082219-015/504, 082119-006/252, 081419-007/135, 081419-005/126, 081419-004/492    REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF TTNS3.00C3-600/UDI (01)000850608002384; REF TTNS3.00C3-900; UDI (01)000850608002391"
    }
  ]
}