{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Billerica",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96617",
      "recalling_firm": "Belmont Instrument LLC",
      "address_1": "780 Boston Rd",
      "address_2": "",
      "postal_code": "01821-5925",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.",
      "recall_number": "Z-1675-2025",
      "product_description": "Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037.      Sterile, single-use, disposable set used with the Hyperthermia Pump.  The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).",
      "product_quantity": "208 sets",
      "reason_for_recall": "Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.",
      "recall_initiation_date": "20250331",
      "center_classification_date": "20250425",
      "report_date": "20250507",
      "code_info": "Model Number: 902-00037. UDI-DI: 00896128002213. Lot Numbers: 20241012, 20241107",
      "more_code_info": ""
    }
  ]
}