  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98434",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: Nationwide Distribution;  International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen;",
      "recall_number": "Z-1673-2026",
      "product_description": "Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;",
      "product_quantity": "298 units (27 US, 271 OUS)",
      "reason_for_recall": "Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.",
      "recall_initiation_date": "20260303",
      "center_classification_date": "20260327",
      "report_date": "20260408",
      "code_info": "(1) System Code: 722063; UDI: 00884838085275; System Serial Number:122, 56, 64, 131, 84, 132, 96, 98, 108, 77, 82, 145, 47, 146, 80, 102, 43, 117, 2, 149, 99, 44, 3, 28, 144, 62, 119, 61, 22, 29, 153, 40, 70, 30, 25, 19, 120, 121, 110, 14, 16, 17, 53, 103, 127, 97, 143, 41, 104, 116, 151, 93, 141, 115, 137, 73, 150, 128, 105, 4, 50, 129, 72, 38, 21, 78, 23, 76, 63, 69, 59, 51, 92, 139, 130, 154, 75, 24, 94, 100, 54, 46, 26, 71, 106, 83, 36, 37, 27, 15, 135, 123, 107, 87, 85, 89, 88, 86, 90, 91, 79, 133, 81, 140, 52, 48, 148, 55, 113;  (2) System Code: 722221; UDI: 00884838085275; System Serial Number:181, 3, 7, 178, 54, 6, 93, 69, 9, 115, 177, 85, 19, 49, 171, 153, 108, 193, 104, 169, 45, 101, 135, 106, 51, 114, 36, 124, 197, 129, 4, 154, 146, 68, 195, 130, 183, 187, 42, 172, 164, 190, 151, 71, 152, 184, 147, 32, 105, 182, 58, 103, 149, 127, 128, 86, 122, 60, 155, 66, 33, 55, 70, 123, 118, 61, 44, 185, 74, 95, 205, 13, 99, 34, 142, 39, 144, 82, 145, 59, 77, 180, 126, 97, 132, 79, 91, 161, 102, 52, 120, 162, 64, 75, 41, 84, 131, 94, 96, 38, 48, 31, 92, 78, 186, 133, 167, 23, 26, 40, 10, 203, 30, 207, 208, 76, 50, 98, 107, 158, 157, 62, 29, 194, 73, 25, 15, 119, 140, 166, 57, 27, 2, 47, 16, 141, 143, 35, 168, 83, 163, 202, 67, 198, 134, 5, 81, 22, 28, 165, 196, 173, 46, 53, 8, 56, 174, 43, 159, 1, 176, 11;  (3) System Code: 722229; UDI: 00884838085275; System Serial Number: 22, 9, 40, 7, 24, 35, 16, 15, 41, 5, 3, 29, 4, 38, 33, 32, 6, 2, 20, 26, 31, 10, 11, 21, 25, 13, 8;",
      "more_code_info": ""
    }
  ]
}