      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85153",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia, Wyoming, and the District of Columbia.",
      "recall_number": "Z-1666-2020",
      "product_description": "BIOMET StageOne Knee Tibial Cement Spacer Mold, 65 MM, Silicone, Sterile, Item 433165.",
      "product_quantity": "230,001",
      "reason_for_recall": "Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.",
      "recall_initiation_date": "20200226",
      "center_classification_date": "20200403",
      "termination_date": "20210729",
      "report_date": "20200415",
      "code_info": "***Lot Numbers***  410520,  959170,  036440,  036450,  036460,  095920,  096370,  318810,  527340  527350,  527360,  528040,  600450,  600530,  688350,  729320,  793180,  857860,  977000,  074660,  074670,  145260,  145270,  145330,  239330,  239340,  239440,  496440,  496450,  496580,  680310,  680320,  749140,  729340,  757950,  821960,  821970,  821980,  926370,  025710,  099750,  165730,  247440,  247460,  247470,  352320,  436890,  436900,  459580,  490080,  490090,  562930,  586710,  632840,  632860,  632870,  708730,  708740,  772900,  969280,  996860,  996900,  010220,  051850,  413420,  413460,  242880,  413430,  413440,  413450,  461670,  518850,  583920,  583930,  667480,  667490,  822040,  056570,  156180,  220730,  270120,  270130,  411640,  411650,  434470,  525320,  525330,  540610,  540620,  634030,  688990,  722580,  738690,  823830,  823880,  823890,  970920,  970930,  070910,  070920,  108360,  207280,  257310,  339530,  413190,  548530,  548540,  588810,  643420,  643450,  672470,  706600,  706610,  907660,  976580,  225810,  225890,  274130,  281770,  281790,  369020,  512310,  512250,  579860,  579870,  625620,  625630,  659470,  659480,  731760,  789370,  789380,  003780,  003790,  270940,  270930,  270950,  281120,  281130,  281150,  281140,  392930,  393010,  393020,  281160,  458250,  535710,  535780,  617630,  617730,  617740,  863660,  863670,  863680,  993800,  993880,  993890,  089310,  089320,  232380,  232390,  958710,  492550,  492560,  492570,  492580,  569610,  739050,  745870,  845120,  900860,  738970,  046350,  981530,  981540,  981470,  212260,  212270,  126180,  126300,  126310,  212150,  361160,  412970,  412980,  361170,  470670,  550190,  753210,  753220,  753230,  913460,  913370,  913470,  131900,  131910,  178000,  226050,  226060,  316520,  316530,  372300,  374920,  374930,  454990,  520680,  611580,  679640,  679650,  776860,  788280,  788290,  046030,  046040,  166200,  166210,  364350,  317650,  317660,  317670,  525950,  525960,  525760,  525850,  697730,  697660,  836040,  923070,  923080,  922970,  047070,  047080,  210260,  210270,  210170,  210120,  351630,  500970,  500980,  500750,  500830,  500850,  644310,  644320,  675070,  806780,  806800,  839260,  955920,  955900,  955830,  081820,  347380,  347370,  347310,  347270,  347280,  486410,  486420,  588210,  588130,  783180,  783120,  903270,  004970,  004980,  005000,  931820,  931870,  088140,  226000,  374530,  432930,  432910,  432920,  432940,  432900,  594300,  891880,  891890,  891910,  891870,  891900,  892140,  892150,  892170,  892210,  892220,  930550,  930560,  930570,  930580,  930590,  930600,  973120,  973130,  973150,  973160,  014720,  014730,  014740,  014750,  022430,  022440,  022460,  022470,  025050,  025130,  025140,  025150,  025160,  026770,  026780,  026930,  026940,  026950,  026960,  115760,  115820,  115860,  202740,  202750,  202760,  202770,  202780,  202790,  202800,  202820,  521440,  521450,  521460,  521470,  521500,  521520,  521560,  521680,  521740,  521750,  521760,  521770,  521780,  521830,  521840,  521850,  521870,  521880,  594310,  528040R,  809210,  177830,  177840,  296680,  325830,  437650,  437660,  618710,  747110,  747310,  747200,  916540,  914340,  916580,  916590,  916650,  916660,  378230,  378290,  378210,  378320,  378330,  378200,  766500,  766620,  766630,  766250,  044070,  044080,  044090,  044120,  044040,  044060,  460840,  025150R,  295620,  295640,  295650,  295670,  295690,  295710,  295720,  295730,  295750,  295760,  321510,  321530,  321570,  321580,  295600,  295630,  295680,  295700,  295740,  295610,  698770,  698790,  698820,  709680,  709710,  711050,  711070,  711090,  711220,  714040,  712230,  714030,  818560,  818570,  844640,  844650,  844660,  845500,  845520,  845540,  851930,  851980,  852000,  879380,  879390,  879410,  880070,  880100,  880110,  894790,  035910,  712190,  894780,  894810,  895300,  915500,  962960,  962990,  963020,  963800,  963810,  963820,  963950,  964750,  964760,  964770,  964900,  995970,  996520,  996540,  035890,  035920,  122890,  122960,  123150,  123160,  123170,  964910,  995830,  996020,  996050,  996490,  996500,  122880,  171110,  171290,  171300,  231790,  231820,  864910,  864920,  864930,  864950,  864960,  806420,  372290,  372350,  484770,  484780,  623170,  623180,  723370,  723380,  723400,  064750,  064740,  089370,  108030,  064970,  108020,  142330,  142350,  237900,  089350,  089360,  197980,  285650,  285680.  ***UDI Numbers***  (01)00880304001824(17)210128(10)410520,  (01)00880304001824(17)221128(10)959170,  (01)00880304001824(17)221203(10)036440,  (01)00880304001824(17)221201(10)036450,  (01)00880304001824(17)221201(10)036460,  (01)00880304001824(17)221211(10)095920,  (01)00880304001824(17)221211(10)096370,  (01)00880304001824(17)230111(10)318810,  (01)00880304001824(17)230118(10)527340,  (01)00880304001824(17)230117(10)527350,  (01)00880304001824(17)230122(10)527360,  (01)00880304001824(17)230118(10)528040,  (01)00880304001824(17)230129(10)600450,  (01)00880304001824(17)230129(10)600530,  (01)00880304001824(17)230211(10)688350,  (01)00880304001824(17)230211(10)729320,  (01)00880304001824(17)230215(10)793180,  (01)00880304001824(17)230228(10)857860,  (01)00880304001824(17)230307(10)977000,  (01)00880304001824(17)230326(10)074660,  (01)00880304001824(17)230326(10)074670,  (01)00880304001824(17)230422(10)145260,  (01)00880304001824(17)230422(10)145270,  (01)00880304001824(17)230422(10)145330,  (01)00880304001824(17)230428(10)239330,  (01)00880304001824(17)230422(10)239340,  (01)00880304001824(17)230422(10)239440,  (01)00880304001824(17)230515(10)496440,  (01)00880304001824(17)230514(10)496450,  (01)00880304001824(17)230515(10)496580,  (01)00880304001824(17)230607(10)680310,  (01)00880304001824(17)230611(10)680320,  (01)00880304001824(17)230612(10)749140,  (01)00880304001824(17)230612(10)729340,  (01)00880304001824(17)230614(10)757950,  (01)00880304001824(17)230625(10)821960,  (01)00880304001824(17)230625(10)821970,  (01)00880304001824(17)230625(10)821980,  (01)00880304001824(17)230702(10)926370,  (01)00880304001824(17)230715(10)025710,  (01)00880304001824(17)230720(10)099750,  (01)00880304001824(17)230726(10)165730,  (01)00880304001824(17)230808(10)247440,  (01)00880304001824(17)230808(10)247460,  (01)00880304001824(17)230808(10)247470,  (01)00880304001824(17)230817(10)352320,  (01)00880304001824(17)230826(10)436890,  (01)00880304001824(17)230826(10)436900,  (01)00880304001824(17)230829(10)459580,  (01)00880304001824(17)230903(10)490080,  (01)00880304001824(17)230903(10)490090,  (01)00880304001824(17)230909(10)562930,  (01)00880304001824(17)230914(10)586710,  (01)00880304001824(17)230914(10)632840,  (01)00880304001824(17)230916(10)632860,  (01)00880304001824(17)230916(10)632870,  (01)00880304001824(17)230925(10)708730,  (01)00880304001824(17)230925(10)708740,  (01)00880304001824(17)230930(10)772900,  (01)00880304001824(17)231021(10)969280,  (01)00880304001824(17)231021(10)996860,  (01)00880304001824(17)231022(10)996900,  (01)00880304001824(17)231025(10)010220,  ,01)00880304001824(17)231025(10)051850,  (01)00880304001824(17)231123(10)413420,  (01)00880304001824(17)231123(10)413460,  (01)00880304001824(17)231201(10)242880,  (01)00880304001824(17)231202(10)413430,  (01)00880304001824(17)231202(10)413440,  (01)00880304001824(17)231202(10)413450,  (01)00880304001824(17)231204(10)461670,  (01)00880304001824(17)231229(10)518850,  (01)00880304001824(17)231212(10)583920,  (01)00880304001824(17)231229(10)583930,  (01)00880304001824(17)240103(10)667480,  (01)00880304001824(17)240103(10)667490,  (01)00880304001824(17)240116(10)822040,  (01)00880304001824(17)240216(10)056570,  (01)00880304001824(17)240217(10)156180,  (01)00880304001824(17)240217(10)220730,  (01)00880304001824(17)240224(10)270120,  (01)00880304001824(17)240222(10)270130,  (01)00880304001824(17)240306(10)411640,  (01)00880304001824(17)240310(10)411650,  (01)00880304001824(17)240309(10)434470,  (01)00880304001824(17)240316(10)525320,  (01)00880304001824(17)240316(10)525330,  (01)00880304001824(17)240316(10)540610,  (01)00880304001824(17)240316(10)540620,  (01)00880304001824(17)240401(10)634030,  (01)00880304001824(17)240402(10)688990,  (01)00880304001824(17)240402(10)722580,  (01)00880304001824(17)240402(10)738690,  (01)00880304001824(17)240408(10)823830,  (01)00880304001824(17)240408(10)823880,  (01)00880304001824(17)240409(10)823890,  (01)00880304001824(17)240421(10)970920,  (01)00880304001824(17)240421(10)970930,  (01)00880304001824(17)240430(10)070910,  (01)00880304001824(17)240429(10)070920,  (01)00880304001824(17)240501(10)108360,  (01)00880304001824(17)240513(10)207280,  (01)00880304001824(17)240519(10)257310,  (01)00880304001824(17)240527(10)339530,  (01)00880304001824(17)240528(10)413190,  (01)00880304001824(17)240610(10)548530,  (01)00880304001824(17)240610(10)548540,  (01)00880304001824(17)240611(10)588810,  (01)00880304001824(17)240618(10)643420,  (01)00880304001824(17)240618(10)643450,  (01)00880304001824(17)240620(10)672470,  (01)00880304001824(17)240708(10)706600,  (01)00880304001824(17)240706(10)706610,  (01)00880304001824(17)240713(10)907660,  (01)00880304001824(17)240720(10)976580,  (01)00880304001824(17)240806(10)225810,  (01)00880304001824(17)240806(10)225890,  (01)00880304001824(17)240813(10)274130,  (01)00880304001824(17)240814(10)281770,  (01)00880304001824(17)240826(10)281790,  (01)00880304001824(17)240820(10)369020,  (01)00880304001824(17)240902(10)512310,  (01)00880304001824(17)240903(10)512250,  (01)00880304001824(17)240909(10)579860,  (01)00880304001824(17)240909(10)579870,  (01)00880304001824(17)240911(10)625620,  (01)00880304001824(17)240912(10)625630,  (01)00880304001824(17)240915(10)659470,  (01)00880304001824(17)240915(10)659480,  (01)00880304001824(17)240919(10)731760,  (01)00880304001824(17)240930(10)789370,  (01)00880304001824(17)240926(10)789380,  (01)00880304001824(17)241020(10)003780,  (01)00880304001824(17)241020(10)003790,  (01)00880304001824(17)241103(10)270940,  (01)00880304001824(17)241107(10)270930,  (01)00880304001824(17)241106(10)270950,  (01)00880304001824(17)241113(10)281120,  (01)00880304001824(17)241120(10)281130,  (01)00880304001824(17)241119(10)281150,  (01)00880304001824(17)241121(10)281140,  (01)00880304001824(17)241201(10)392930,  (01)00880304001824(17)241201(10)393010,  (01)00880304001824(17)241130(10)393020,  (01)00880304001824(17)241208(10)281160,  (01)00880304001824(17)241208(10)458250,  (01)00880304001824(17)241209(10)535710,  (01)00880304001824(17)241209(10)535780,  (01)00880304001824(17)241222(10)617630,  (01)00880304001824(17)241216(10)617730,  (01)00880304001824(17)241215(10)617740,  (01)00880304001824(17)250119(10)863660,  (01)00880304001824(17)250119(10)863670,  (01)00880304001824(17)250114(10)863680,  (01)00880304001824(17)250128(10)993800,  (01)00880304001824(17)250122(10)993880,  (01)00880304001824(17)250122(10)993890,  (01)00880304001824(17)250202(10)089310,  (01)00880304001824(17)250202(10)089320,  (01)00880304001824(17)250219(10)232380,  (01)00880304001824(17)250225(10)232390,  (01)00880304001824(17)250311(10)958710,  (01)00880304001824(17)250324(10)492550,  (01)00880304001824(17)250324(10)492560,  (01)00880304001824(17)250324(10)492570,  (01)00880304001824(17)250324(10)492580,  (01)00880304001824(17)250409(10)569610,  (01)00880304001824(17)250414(10)739050,  (01)00880304001824(17)250427(10)745870,  (01)00880304001824(17)250427(10)845120,  (01)00880304001824(17)250424(10)900860,  (01)00880304001824(17)250428(10)738970,  (01)00880304001824(17)250505(10)046350,  (01)00880304001824(17)250430(10)981530,  (01)00880304001824(17)250501(10)981540,  (01)00880304001824(17)250507(10)981470,  (01)00880304001824(17)250527(10)212260,  (01)00880304001824(17)250527(10)212270,  (01)00880304001824(17)250528(10)126180,  (01)00880304001824(17)250528(10)126300,  (01)00880304001824(17)250601(10)126310,  (01)00880304001824(17)250611(10)212150,  (01)00880304001824(17)250616(10)361160,  (01)00880304001824(17)250612(10)412970,  (01)00880304001824(17)250612(10)412980,  (01)00880304001824(17)250618(10)361170,  (01)00880304001824(17)250618(10)470670,  (01)00880304001824(17)250625(10)550190,  (01)00880304001824(17)250717(10)753210,  (01)00880304001824(17)250716(10)753220,  (01)00880304001824(17)250717(10)753230,  (01)00880304001824(17)250804(10)913460,  (01)00880304001824(17)250810(10)913370,  (01)00880304001824(17)250805(10)913470,  (01)00880304001824(17)250822(10)131900,  (01)00880304001824(17)250821(10)131910,  (01)00880304001824(17)250901(10)178000,  (01)00880304001824(17)250908(10)226050,  (01)00880304001824(17)250910(10)226060,  (01)00880304001824(17)250910(10)316520,  (01)00880304001824(17)250910(10)316530,  (01)00880304001824(17)250923(10)372300,  (01)00880304001824(17)250923(10)374920,  (01)00880304001824(17)250923(10)374930,  (01)00880304001824(17)250928(10)454990,  (01)00880304001824(17)250928(10)520680,  (01)00880304001824(17)251012(10)611580,  (01)00880304001824(17)251012(10)679640,  (01)00880304001824(17)251013(10)679650,  (01)00880304001824(17)251023(10)776860,  (01)00880304001824(17)251104(10)788280,  (01)00880304001824(17)251104(10)788290,  (01)00880304001824(17)251117(10)046030,  (01)00880304001824(17)251117(10)046040,  (01)00880304001824(17)251202(10)166200,  (01)00880304001824(17)251202(10)166210,  (01)00880304001824(17)251211(10)364350,  (01)00880304001824(17)251218(10)317650,  (01)00880304001824(17)251218(10)317660,  (01)00880304001824(17)251219(10)317670,  (01)00880304001824(17)251229(10)525950,  (01)00880304001824(17)251230(10)525960,  (01)00880304001824(17)260112(10)525760,  (01)00880304001824(17)260108(10)525850,  (01)00880304001824(17)260113(10)697730,  (01)00880304001824(17)260118(10)697660,  (01)00880304001824(17)260125(10)836040,  (01)00880304001824(17)260122(10)836050,  (01)00880304001824(17)260203(10)923070,  (01)00880304001824(17)260203(10)923080,  (01)00880304001824(17)260209(10)922970,  (01)00880304001824(17)260224(10)047070,  (01)00880304001824(17)260225(10)047080,  (01)00880304001824(17)260228(10)210260,  (01)00880304001824(17)260226(10)210270,  (01)00880304001824(17)260304(10)210170,  (01)00880304001824(17)260310(10)210120,  (01)00880304001824(17)260314(10)351630,  (01)00880304001824(17)260322(10)500970,  (01)00880304001824(17)260322(10)500980,  (01)00880304001824(17)260329(10)500750,  (01)00880304001824(17)260325(10)500830,  (01)00880304001824(17)260325(10)500850,  (01)00880304001824(17)260404(10)644310,  (01)00880304001824(17)260404(10)644320,  (01)00880304001824(17)260406(10)675070,  (01)00880304001824(17)260419(10)806780,  (01)00880304001824(17)260418(10)806800,  (01)00880304001824(17)260420(10)839260,  (01)00880304001824(17)260427(10)955920,  (01)00880304001824(17)260429(10)955900,  (01)00880304001824(17)260504(10)955830,  (01)00880304001824(17)260512(10)081820,  (01)00880304001824(17)260601(10)347380,  (01)00880304001824(17)260607(10)347370,  (01)00880304001824(17)260608(10)347310,  (01)00880304001824(17)260610(10)347270,  (01)00880304001824(17)260610(10)347280,  (01)00880304001824(17)260617(10)486410,  (01)00880304001824(17)260614(10)486420,  (01)00880304001824(17)260627(10)588210,  (01)00880304001824(17)260630(10)588130,  (01)00880304001824(17)260708(10)783180,  (01)00880304001824(17)260714(10)783120,  (01)00880304001824(17)260718(10)903270,  (01)00880304001824(17)260726(10)004970,  (01)00880304001824(17)260725(10)004980,  (01)00880304001824(17)260725(10)005000,  (01)00880304001824(17)260722(10)931820,  (01)00880304001824(17)260722(10)931870,  (01)00880304001824(17)260804(10)088140,  (01)00880304001824(17)260815(10)226000,  (01)00880304001824(17)260827(10)374530,  (01)00880304001824(17)260827(10)432930,  (01)00880304001824(17)260831(10)432910,  (01)00880304001824(17)260901(10)432920,  (01)00880304001824(17)260831(10)432940,  (01)00880304001824(17)260909(10)432900,  (01)00880304001824(17)260927(10)594300,  (01)00880304001824(17)270609(10)891880,  (01)00880304001824(17)270609(10)891890,  (01)00880304001824(17)270609(10)891910,  (01)00880304001824(17)270616(10)891870,  (01)00880304001824(17)270615(10)891900,  (01)00880304001824(17)270616(10)892140,  (01)00880304001824(17)270615(10)892150,  (01)00880304001824(17)270609(10)892170,  (01)00880304001824(17)270615(10)892210,  (01)00880304001824(17)270615(10)892220,  (01)00880304001824(17)270616(10)930550,  (01)00880304001824(17)270616(10)930560,  (01)00880304001824(17)270616(10)930570,  (01)00880304001824(17)270616(10)930580,  (01)00880304001824(17)270616(10)930590,  (01)00880304001824(17)270616(10)930600,  (01)00880304001824(17)270615(10)973120,  (01)00880304001824(17)270616(10)973130,  (01)00880304001824(17)270615(10)973150,  (01)00880304001824(17)270616(10)973160,  (01)00880304001824(17)270623(10)014720,  (01)00880304001824(17)270623(10)014730,  (01)00880304001824(17)270624(10)014740,  (01)00880304001824(17)270622(10)014750,  (01)00880304001824(17)270622(10)022430,  (01)00880304001824(17)270623(10)022440,  (01)00880304001824(17)270622(10)022460,  (01)00880304001824(17)270622(10)022470,  (01)00880304001824(17)270622(10)025050,  (01)00880304001824(17)270622(10)025130,  (01)00880304001824(17)270623(10)025140,  (01)00880304001824(17)270623(10)025150,  (01)00880304001824(17)270622(10)025160,  (01)00880304001824(17)270622(10)026770,  (01)00880304001824(17)270623(10)026780,  (01)00880304001824(17)270623(10)026930,  (01)00880304001824(17)270622(10)026940,  (01)00880304001824(17)270623(10)026950,  (01)00880304001824(17)270623(10)026960,  (01)00880304001824(17)270625(10)115760,  (01)00880304001824(17)270627(10)115820,  (01)00880304001824(17)270626(10)115860,  (01)00880304001824(17)270630(10)202740,  (01)00880304001824(17)270630(10)202750,  (01)00880304001824(17)270630(10)202760,  (01)00880304001824(17)270630(10)202770,  (01)00880304001824(17)270629(10)202780,  (01)00880304001824(17)270630(10)202790,  (01)00880304001824(17)270629(10)202800,  (01)00880304001824(17)270629(10)202820,  (01)00880304001824(17)270725(10)521440,  (01)00880304001824(17)270726(10)521450,  (01)00880304001824(17)270726(10)521460,  (01)00880304001824(17)270725(10)521470,  (01)00880304001824(17)270725(10)521500,  (01)00880304001824(17)270726(10)521520,  (01)00880304001824(17)270726(10)521560,  (01)00880304001824(17)270725(10)521680,  (01)00880304001824(17)270726(10)521740,  (01)00880304001824(17)270726(10)521750,  (01)00880304001824(17)270726(10)521760,  (01)00880304001824(17)270725(10)521770,  (01)00880304001824(17)270726(10)521780,  (01)00880304001824(17)270726(10)521830,  (01)00880304001824(17)270726(10)521840,  (01)00880304001824(17)270726(10)521850,  (01)00880304001824(17)270725(10)521870,  (01)00880304001824(17)270726(10)521880,  (01)00880304001824(17)270824(10)594310,  (01)00880304001824(17)230118(10)528040R,  (01)00880304001824(17)271111(10)809210,  (01)00880304001824(17)271127(10)177830,  (01)00880304001824(17)271128(10)177840,  (01)00880304001824(17)271210(10)296680,  (01)00880304001824(17)271217(10)325830,  (01)00880304001824(17)271221(10)437650,  (01)00880304001824(17)271220(10)437660,  (01)00880304001824(17)280105(10)618710,  (01)00880304001824(17)280117(10)747110,  (01)00880304001824(17)280117(10)747310,  (01)00880304001824(17)280119(10)747200,  (01)00880304001824(17)280131(10)916540,  (01)00880304001824(17)280201(10)914340,  (01)00880304001824(17)280201(10)916580,  (01)00880304001824(17)280201(10)916590,  (01)00880304001824(17)280201(10)916650,  (01)00880304001824(17)280201(10)916660,  (01)00880304001824(17)280228(10)378230,  (01)00880304001824(17)280228(10)378290,  (01)00880304001824(17)280306(10)378210,  (01)00880304001824(17)280306(10)378320,  (01)00880304001824(17)280228(10)378330,  (01)00880304001824(17)280305(10)378200,  (01)00880304001824(17)280322(10)766500,  (01)00880304001824(17)280403(10)766620,  (01)00880304001824(17)280330(10)766630,  (01)00880304001824(17)280405(10)766250,  (01)00880304001824(17)280417(10)044070,  (01)00880304001824(17)280417(10)044080,  (01)00880304001824(17)280417(10)044090,  (01)00880304001824(17)280417(10)044120,  (01)00880304001824(17)280502(10)044040,  (01)00880304001824(17)280503(10)044060,  (01)00880304001824(17)280509(10)460840,  (01)00880304001824(17)270623(10)025150R,  (01)00880304001824(17)290126(10)295620,  (01)00880304001824(17)290126(10)295640,  (01)00880304001824(17)290131(10)295650,  (01)00880304001824(17)290126(10)295670,  (01)00880304001824(17)290129(10)295690,  (01)00880304001824(17)290126(10)295710,  (01)00880304001824(17)290129(10)295720,  (01)00880304001824(17)290126(10)295730,  (01)00880304001824(17)290201(10)295750,  (01)00880304001824(17)290127(10)295760,  (01)00880304001824(17)290205(10)321510,  (01)00880304001824(17)290204(10)321530,  (01)00880304001824(17)290204(10)321570,  (01)00880304001824(17)290204(10)321580,  (01)00880304001824(17)290127(10)295600,  (01)00880304001824(17)290128(10)295630,  (01)00880304001824(17)290127(10)295680,  (01)00880304001824(17)290129(10)295700,  (01)00880304001824(17)290126(10)295740,  (01)00880304001824(17)290221(10)295610,  (01)00880304001824(17)290228(10)698770,  (01)00880304001824(17)290228(10)698790,  (01)00880304001824(17)290228(10)698820,  (01)00880304001824(17)290306(10)709680,  (01)00880304001824(17)290305(10)709710,  (01)00880304001824(17)290228(10)711050,  (01)00880304001824(17)290228(10)711070,  (01)00880304001824(17)290301(10)711090,  (01)00880304001824(17)290304(10)711220,  (01)00880304001824(17)290305(10)714040,  (01)00880304001824(17)290309(10)712230,  (01)00880304001824(17)290311(10)714030,  (01)00880304001824(17)290311(10)818560,  (01)00880304001824(17)290310(10)818570,  (01)00880304001824(17)290307(10)844640,  (01)00880304001824(17)290307(10)844650,  (01)00880304001824(17)290306(10)844660,  (01)00880304001824(17)290310(10)845500,  (01)00880304001824(17)290306(10)845520,  (01)00880304001824(17)290308(10)845540,  (01)00880304001824(17)290308(10)851930,  (01)00880304001824(17)290308(10)851980,  (01)00880304001824(17)290308(10)852000,  (01)00880304001824(17)290311(10)879380,  (01)00880304001824(17)290312(10)879390,  (01)00880304001824(17)290312(10)879410,  (01)00880304001824(17)290313(10)880070,  (01)00880304001824(17)290312(10)880100,  (01)00880304001824(17)290312(10)880110,  (01)00880304001824(17)290313(10)894790,  (01)00880304001824(17)290319(10)035910,  (01)00880304001824(17)290315(10)712190,  (01)00880304001824(17)290314(10)894780,  (01)00880304001824(17)290313(10)894810,  (01)00880304001824(17)290316(10)895300,  (01)00880304001824(17)290315(10)915500,  (01)00880304001824(17)290316(10)962960,  (01)00880304001824(17)290315(10)962990,  (01)00880304001824(17)290316(10)963020,  (01)00880304001824(17)290316(10)963800,  (01)00880304001824(17)290316(10)963810,  (01)00880304001824(17)290314(10)963820,  (01)00880304001824(17)290315(10)963950,  (01)00880304001824(17)290314(10)964750,  (01)00880304001824(17)290314(10)964760,  (01)00880304001824(17)290315(10)964770,  (01)00880304001824(17)290315(10)964900,  (01)00880304001824(17)290319(10)995970,  (01)00880304001824(17)290319(10)996520,  (01)00880304001824(17)290319(10)996540,  (01)00880304001824(17)290321(10)035890,  (01)00880304001824(17)290320(10)035920,  (01)00880304001824(17)290322(10)122890,  (01)00880304001824(17)290322(10)122960,  (01)00880304001824(17)290322(10)123150,  (01)00880304001824(17)290322(10)123160,  (01)00880304001824(17)290322(10)123170,  (01)00880304001824(17)290320(10)964910,  (01)00880304001824(17)290320(10)995830,  (01)00880304001824(17)290321(10)996020,  (01)00880304001824(17)290322(10)996050,  (01)00880304001824(17)290320(10)996490,  (01)00880304001824(17)290321(10)996500,  (01)00880304001824(17)290325(10)122880,  (01)00880304001824(17)290329(10)171110,  (01)00880304001824(17)290328(10)171290,  (01)00880304001824(17)290329(10)171300,  (01)00880304001824(17)290327(10)231790,  (01)00880304001824(17)290327(10)231820,  (01)00880304001824(17)290511(10)864910,  (01)00880304001824(17)290510(10)864920,  (01)00880304001824(17)290510(10)864930,  (01)00880304001824(17)290510(10)864950,  (01)00880304001824(17)290510(10)864960,  (01)00880304001824(17)290925(10)806420,  (01)00880304001824(17)291024(10)372290,  (01)00880304001824(17)291024(10)372350,  (01)00880304001824(17)291024(10)484770,  (01)00880304001824(17)291024(10)484780,  (01)00880304001824(17)291111(10)623170,  (01)00880304001824(17)291110(10)623180,  (01)00880304001824(17)291107(10)723370,  (01)00880304001824(17)291107(10)723380,  (01)00880304001824(17)291107(10)723400,  (01)00880304001824(17)291211(10)064750,  (01)00880304001824(17)291212(10)064740,  (01)00880304001824(17)291220(10)089370,  (01)00880304001824(17)291220(10)108030,  (01)00880304001824(17)300103(10)064970,  (01)00880304001824(17)300103(10)108020,  (01)00880304001824(17)300111(10)142330,  (01)00880304001824(17)300113(10)142350,  (01)00880304001824(17)300113(10)237900,  (01)00880304001824(17)300118(10)089350,  (01)00880304001824(17)300122(10)089360,  (01)00880304001824(17)300123(10)197980,  (01)00880304001824(17)300122(10)285650,  (01)00880304001824(17)300124(10)285680"
    }
  ]
}