      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73667",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr Ms 514",
      "address_2": "PO BOX 6101",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands",
      "recall_number": "Z-1663-2016",
      "product_description": "Dimension Vista  Assays: Direct HDL Cholesterol (AHDL)",
      "product_quantity": "137422 units total",
      "reason_for_recall": "Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays.  Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole.  NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific",
      "recall_initiation_date": "20160317",
      "center_classification_date": "20160511",
      "termination_date": "20170925",
      "report_date": "20160518",
      "code_info": "Device Listing: D056475 Direct HDL Cholesterol (TRIG) Catalog # K3048A"
    }
  ]
}