  "results": [
    {
      "status": "Ongoing",
      "city": "Drachten",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98473",
      "recalling_firm": "BD KIESTRA LAB AUTOMATION",
      "address_1": "Marconilaan 6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: KS, MD, MN, NJ, TX, VA;  International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;",
      "recall_number": "Z-1658-2026",
      "product_description": "BD Kiestra\" ReadA;  Catalog No.: 446948.",
      "product_quantity": "82 units (7 US, 65 OUS)",
      "reason_for_recall": "In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.",
      "recall_initiation_date": "20260206",
      "center_classification_date": "20260327",
      "report_date": "20260408",
      "code_info": "Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All;",
      "more_code_info": ""
    }
  ]
}