{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Abbott Park",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96567",
      "recalling_firm": "Abbott Laboratories",
      "address_1": "100 Abbott Park Rd Bldg Ap8b",
      "address_2": "N/A",
      "postal_code": "60064-3502",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.",
      "recall_number": "Z-1652-2025",
      "product_description": "Brand Name: CELL-DYN Ruby  Product Name: CELL-DYN Ruby  Model/Catalog Number: 08H67-01  Software Version: All software versions  Product Description: CELL-DYN Ruby  Component: No",
      "product_quantity": "2947",
      "reason_for_recall": "When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.",
      "recall_initiation_date": "20250318",
      "center_classification_date": "20250422",
      "report_date": "20250430",
      "code_info": "UDI-DI 00380740017170  SNs: 30039BG 34021BG 34079BG 34108BG 34123BG 34138BG 34140BG 34149BG 34152BG 34161BG 34177BG 34186BG 34232BG 34242BG 34272BG 34288BG 34302BG 34323BG 34339BG 34349BG 34353BG 34354BG 34364BG 34372BG 34391BG 34395BG 34414BG 34433BG 34439BG 34448BG 34451BG 34458BG 34466BG 34468BG 34473BG 34478BG 34488BG 34491BG 34531BG 34533BG 34539BG 34551BG 34555BG 34559BG 34565BG 34566BG 34598BG 34604BG 34605BG 34620BG 34633BG 34634BG 34646BG 34667BG 34669BG 34695BG 34726BG 34734BG 34736BG 34742BG 34744BG 34758BG 34768BG 34788BG 34789BG 34791BG 34793BG 34805BG 34816BG 34837BG 34839BG 34841BG 34844BG 34850BG 34868BG 34878BG 34881BG 34884BG 34888BG 34893BG 34902BG 34903BG 34907BG 34918BG 34927BG 34933BG 34940BG 34942BG 34950BG 34953BG 34957BG 34965BG 34971BG 34974BG 34975BG 34976BG 34980BG 34984BG 34988BG 34997BG 35016BG 35029BG 35045BG 35049BG 35050BG 35056BG 35061BG 35062BG 35066BG 35067BG 35073BG 35077BG 35089BG 35097BG 35102BG 35104BG 35105BG 35111BG 35112BG 35117BG 35127BG 35128BG 35130BG 35135BG 35142BG 35143BG 35148BG 35154BG 35155BG 35161BG 35163BG 35165BG 35167BG 35168BG 35172BG 35173BG 35174BG 35175BG 35176BG 35177BG 35178BG 35179BG 35180BG 35181BG 35182BG 35184BG 35191BG 35192BG 35201BG 35207BG 35217BG 35221BG 35224BG 35253BG 35259BG 35272BG 35273BG 35274BG 35275BG 35276BG 35277BG 35278BG 35279BG 35280BG 35281BG 35285BG 35295BG 35300BG 35304BG 35305BG 35305BG 35307BG 35309BG 35311BG 35317BG 35320BG 35324BG 35334BG 35338BG 35341BG 35344BG 35346BG 35355BG 35370BG 35372BG 35376BG 35377BG 35380BG 35381BG 35383BG 35389BG 35391BG 35396BG 35397BG 35398BG 35401BG 35404BG 35413BG 35417BG 35420BG 35424BG 35439BG 35442BG 35443BG 35449BG 35451BG 35457BG 35470BG 35471BG 35478BG 35480BG 35496BG 35500BG 35507BG 35509BG 35511BG 35513BG 35522BG 35524BG 35525BG 35529BG 35531BG 35534BG 35535BG 35548BG 35550BG 35554BG 35555BG 35556BG 35558BG 35561BG 35563BG 35564BG 35573BG 35578BG 35585BG 35594BG 35597BG 35598BG 35599BG 35600BG 35601BG 35605BG 35626BG 35677BG 35682BG 35688BG 35689BG 35692BG 35694BG 35699BG 35700BG 35701BG 35717BG 35718BG 35730BG 35740BG 35751BG 35752BG 35755BG 35758BG 35760BG 35761BG 35762BG 35763BG 35767BG 35772BG 35778BG 35779BG 35780BG 35785BG 35788BG 35789BG 35796BG 35798BG 35803BG 35804BG 35809BG 35811BG 35819BG 35820BG 35822BG 35825BG 35826BG 35830BG 35835BG 35836BG 35837BG 35839BG 35841BG 35842BG 35844BG 35846BG 35847BG 35848BG 35850BG 35856BG 35858BG 35871BG 35877BG 35878BG 35879BG 35881BG 35891BG 35892BG 35893BG 35905BG 35950BG 35952BG 35953BG 35959BG 35960BG 35962BG 35965BG 35966BG 35969BG 35973BG 35974BG 35976BG 35987BG 35990BG 35993BG 35995BG 36000BG 36002BG 36004BG 36007BG 36008BG 36009BG 36011BG 36012BG 36020BG 36023BG 36024BG 36025BG 36026BG 36029BG 36032BG 36037BG 36042BG 36044BG 36052BG 36055BG 36057BG 36059BG 36060BG 36061BG 36062BG 36063BG 36068BG 36069BG 36072BG 36073BG 36074BG 36075BG 36079BG 36082BG 36089BG 36091BG 36092BG 36093BG 36095BG 36101BG 36102BG 36104BG 36105BG 36111BG 36116BG 36117BG 36119BG 36120BG 36123BG 36124BG 36125BG 36127BG 36130BG 36139BG 36153BG 36155BG 36166BG 36175BG 36178BG 36188BG 36196BG 36197BG 36199BG 36214BG 36216BG 36219BG 36224BG 36230BG 36231BG 36237BG 36249BG 36250BG 36253BG 36256BG 36273BG 36274BG 36275BG 36276BG 36279BG 36282BG 36283BG 36285BG 36287BG 36288BG 36309BG 36310BG 36313BG 36314BG 36317BG 36318BG 36319BG 36320BG 36330BG 36335BG 36339BG 36349BG 36354BG 36357BG 36362BG 36363BG 36367BG 36368BG 36369BG 36370BG 36371BG 36375BG 36377BG 36378BG 36379BG 36380BG 36382BG 36383BG 36384BG 36385BG 36387BG 36388BG 36391BG 36392BG 36394BG 36398BG 36400BG 36401BG 36402BG 36407BG 36408BG 36410BG 36418BG 36424BG 36426BG 36427BG 36428BG 36429BG 36435BG 36437BG 36441BG 36442BG 36447BG 36448BG 36454BG 36455BG 36456BG 36462BG 36464BG 36465BG 36466BG 36467BG 36468BG 36469BG 36471BG 36473BG 36474BG 36478BG 36483BG 36485BG 36487BG 36488BG 36489BG 36497BG 36501BG 36505BG 36507BG 36526BG 36531BG 36537BG 36538BG 36540BG 36541BG 36552BG 36564BG 36566BG 36567BG 36569BG 36570BG 36571BG 36572BG 36584BG 36586BG 36590BG 36591BG 36597BG 36602BG 36604BG 36606BG 36610BG 36624BG 36625BG 36633BG 36634BG 36636BG 36638BG 36641BG 36643BG 36644BG 36647BG 36648BG 36649BG 36651BG 36653BG 36656BG 36660BG 36671BG 36672BG 36678BG 36682BG 36694BG 36695BG 36697BG 36699BG 36700BG 36702BG 36703BG 36704BG 36709BG 36710BG 36714BG 36720BG 36724BG 36725BG 36726BG 36728BG 36729BG 36730BG 36731BG 36732BG 36734BG 36735BG 36739BG 36740BG 36741BG 36748BG 36750BG 36753BG 36756BG 36764BG 36775BG 36781BG 36784BG 36790BG 36798BG 36799BG 36801BG 36804BG 36805BG 36808BG 36810BG 36811BG 36812BG 36813BG 36816BG 36817BG 36822BG 36828BG 36829BG 36836BG 36837BG 36838BG 36839BG 36846BG 36868BG 36869BG 36873BG 36875BG 36881BG 36884BG 36886BG 36888BG 36889BG 36890BG 36891BG 36892BG 36893BG 36895BG 36896BG 36897BG 36900BG 36903BG 36907BG 36909BG 36913BG 36914BG 36915BG 54003BG 54004BG 54005BG 54006BG 54007BG 54008BG 54009BG 54013BG 54015BG 54021BG 54023BG 54036BG 54039BG 54040BG 54041BG 54049BG 54052BG 54057BG 54060BG 54063BG 54071BG 54074BG 54084BG 54093BG 54094BG 54098BG 54100BG 54102BG 54103BG 54104BG 54107BG 54108BG 54110BG 54115BG 54117BG 54121BG 54122BG 54123BG 54124BG 54125BG 54126BG 54128BG 54129BG 54135BG 54140BG 54142BG 54143BG 54151BG 54154BG 54171BG 54172BG 54178BG 54180BG 54181BG 54182BG 54186BG 54187BG 54189BG 54190BG 54193BG 54196BG 54207BG 54208BG 54209BG 54211BG 54212BG 54215BG 54216BG 54218BG 54222BG 54226BG 54230BG 54231BG 54232BG 54234BG 54235BG 54241BG 54243BG 54248BG 54255BG 54256BG 54257BG 54258BG 54262BG 54263BG 54264BG 54268BG 54270BG 54271BG 54272BG 54273BG 54278BG 54286BG 54287BG 54295BG 54296BG 54297BG 54299BG 54300BG 54301BG 54307BG 54308BG 54311BG 54316BG 54320BG 54325BG 54326BG 54329BG 54330BG 54331BG 54332BG 54333BG 54334BG 54336BG 54338BG 54339BG 54341BG 54343BG 54345BG 54347BG 54349BG 54350BG 54351BG 54355BG 54356BG 54360BG 54362BG 54371BG 54372BG 54373BG 54374BG 54376BG 54383BG 54384BG 54394BG 54395BG 54397BG 54399BG 54402BG 54405BG 54409BG 54411BG 54412BG 54414BG 54416BG 54421BG 54425BG 54434BG 54435BG 54437BG 54439BG 54442BG 54444BG 54445BG 54447BG 54456BG 54457BG 54460BG 54462BG 54463BG 54464BG 54468BG 54469BG 54470BG 54471BG 54474BG 54475BG 54476BG 54477BG 54480BG 54481BG 54489BG 54492BG 54496BG 54499BG 54500BG 54502BG 54504BG 54505BG 54508BG 54509BG 54510BG 54511BG 54513BG 54514BG 54515BG 54518BG 54520BG 54525BG 54527BG 54528BG 54529BG 54530BG 54532BG 54536BG 54537BG 54540BG 54546BG 54550BG 54556BG 54557BG 54558BG 54560BG 54561BG 54563BG 54564BG 54566BG 54567BG 54578BG 54581BG 54582BG 54588BG 54590BG 54593BG 54594BG 54595BG 54600BG 54605BG 54608BG 54609BG 54610BG 54611BG 54612BG 54613BG 54616BG 54617BG 54619BG 54621BG 54625BG 54626BG 54627BG 54632BG 54634BG 54636BG 54637BG 54638BG 54639BG 54640BG 54644BG 54645BG 54647BG 54648BG 54650BG 54655BG 54656BG 54657BG 54659BG 54666BG 54669BG 54672BG 54677BG 54679BG 54680BG 54681BG 54682BG 54683BG 54685BG 54687BG 54689BG 54690BG 54694BG 54700BG 54705BG 54706BG 54707BG 54708BG 54711BG 54712BG 54713BG 54714BG 54715BG 54716BG 54718BG 54719BG 54721BG 54722BG 54723BG 54724BG 54732BG 54736BG 54738BG 54739BG 54740BG 54741BG 54742BG 54743BG 54746BG 54748BG 54749BG 54750BG 54754BG 54757BG 54760BG 54762BG 54763BG 54764BG 54768BG 54769BG 54770BG 54773BG 54774BG 54775BG 54778BG 54779BG 54781BG 54783BG 54786BG 54788BG 54790BG 54793BG 54796BG 54798BG 54799BG 54807BG 54809BG 54810BG 54812BG 54814BG 54817BG 54819BG 54821BG 54822BG 54823BG 54824BG 54829BG 54831BG 54837BG 54839BG 54841BG 54843BG 54845BG 54846BG 54848BG 54852BG 54857BG 54860BG 54863BG 54866BG 54867BG 54868BG 54870BG 54873BG 54874BG 54878BG 54879BG 54881BG 54885BG 54887BG 54892BG 54895BG 54899BG 54902BG 54903BG 54905BG 54907BG 54908BG 54909BG 54910BG 54912BG 54913BG 54920BG 54925BG 54926BG 54928BG 54929BG 54930BG 54931BG 54934BG 54935BG 54936BG 54937BG 54939BG 54941BG 54945BG 54946BG 54948BG 54949BG 54951BG 54954BG 54955BG 54958BG 54960BG 54963BG 54964BG 54965BG 54966BG 54968BG 54969BG 54970BG 54977BG 54978BG 54979BG 54980BG 54983BG 54984BG 54986BG 54992BG 54993BG 54996BG 55000BG 55004BG 55006BG 55009BG 55011BG 55014BG 55015BG 55018BG 55025BG 55027BG 55030BG 55031BG 55032BG 55033BG 55034BG 55035BG 55037BG 55038BG 55039BG 55041BG 55047BG 55048BG 55051BG 55063BG 55065BG 55066BG 55067BG 55068BG 55070BG 55071BG 55073BG 55074BG 55078BG 55080BG 55084BG 55087BG 55095BG 55099BG 55100BG 55101BG 55103BG 55104BG 55105BG 55107BG 55108BG 55113BG 55116BG 55117BG 55118BG 55124BG 55127BG 55133BG 55137BG 55138BG 55139BG 55144BG 55147BG 55148BG 55149BG 55150BG 55151BG 55153BG 55156BG 55158BG 55159BG 55162BG 55163BG 55164BG 55165BG 55166BG 55167BG 55168BG 55169BG 55170BG 55171BG 55172BG 55174BG 55177BG 55178BG 55182BG 55184BG 55185BG 55186BG 55187BG 55190BG 55192BG 55196BG 55197BG 55199BG 55210BG 55211BG 55217BG 55218BG 55220BG 55221BG 55222BG 55224BG 55225BG 55226BG 55228BG 55231BG 55232BG 55237BG 55241BG 55242BG 55243BG 55244BG 55245BG 55249BG 55255BG 55256BG 55258BG 55259BG 55260BG 55261BG 55265BG 55266BG 55267BG 55268BG 55269BG 55270BG 55272BG 55273BG 55274BG 55280BG 55281BG 55285BG 55289BG 55290BG 55291BG 55294BG 55295BG 55297BG 55301BG 55302BG 55305BG 55307BG 55310BG 55313BG 55314BG 55315BG 55316BG 55317BG 55319BG 55320BG 55322BG 55325BG 55329BG 55330BG 55331BG 55333BG 55335BG 55336BG 55337BG 55343BG 55344BG 55345BG 55350BG 55355BG 55358BG 55359BG 55360BG 55368BG 55374BG 55391BG 55395BG 55400BG 55403BG 55404BG 55405BG 55406BG 55410BG 55412BG 55415BG 55416BG 55418BG 55421BG 55427BG 55430BG 55440BG 55441BG 55444BG 55445BG 55447BG 55449BG 55450BG 55453BG 55455BG 55459BG 55460BG 55462BG 55463BG 55464BG 55465BG 55466BG 55470BG 55471BG 55472BG 55478BG 55481BG 55487BG 55489BG 55490BG 55491BG 55495BG 55499BG 55500BG 55501BG 55502BG 55504BG 55506BG 55509BG 55510BG 55511BG 55514BG 55515BG 55516BG 55522BG 55523BG 55524BG 55526BG 55527BG 55528BG 55530BG 55531BG 55534BG 55536BG 55537BG 55539BG 55541BG 55543BG 55544BG 55547BG 55548BG 55549BG 55551BG 55552BG 55553BG 55554BG 55555BG 55556BG 55557BG 55560BG 55561BG 55562BG 55563BG 55564BG 55566BG 55567BG 55573BG 55574BG 55581BG 55582BG 55583BG 55584BG 55593BG 55595BG 55597BG 55598BG 55599BG 55600BG 55601BG 55602BG 55603BG 55607BG 55608BG 55611BG 55612BG 55613BG 55614BG 55615BG 55617BG 55618BG 55619BG 55620BG 55621BG 55622BG 55623BG 55624BG 55629BG 55630BG 55631BG 55632BG 55637BG 55638BG 55639BG 55640BG 55641BG 55642BG 55643BG 55645BG 55646BG 55647BG 55648BG 55649BG 55653BG 55654BG 55655BG 55658BG 55660BG 55661BG 55663BG 55666BG 55667BG 55669BG 55670BG 55678BG 55679BG 55680BG 55681BG 55682BG 55683BG 55687BG 55690BG 55692BG 55693BG 55694BG 55698BG 55699BG 55700BG 55701BG 55705BG 55709BG 55710BG 55711BG 55712BG 55713BG 55715BG 55716BG 55717BG 55718BG 55721BG 55724BG 55725BG 55726BG 55728BG 55732BG 55733BG 55735BG 55736BG 55737BG 55738BG 55740BG 55741BG 55742BG 55743BG 55744BG 55745BG 55749BG 55752BG 55753BG 55754BG 55758BG 55760BG 55761BG 55763BG 55764BG 55767BG 55768BG 55770BG 55771BG 55772BG 55775BG 55778BG 55779BG 55780BG 55784BG 55789BG 55792BG 55795BG 55798BG 55800BG 55804BG 55806BG 55807BG 55809BG 55811BG 55812BG 55816BG 55819BG 55821BG 55822BG 55823BG 55824BG 55825BG 55826BG 55827BG 55829BG 55831BG 55833BG 55835BG 55837BG 55840BG 55841BG 55842BG 55843BG 55844BG 55845BG 55846BG 55848BG 55849BG 55850BG 55851BG 55852BG 55855BG 55856BG 55857BG 55858BG 55862BG 55863BG 55864BG 55865BG 55867BG 55875BG 55878BG 55880BG 55882BG 55891BG 55893BG 55894BG 55898BG 55900BG 55901BG 55902BG 55904BG 70002BG 70003BG 70006BG 70007BG 70008BG 70009BG 70010BG 70013BG 70016BG 70018BG 70020BG 70022BG 70024BG 70025BG 70026BG 70027BG 70030BG 70031BG 70033BG 70035BG 70038BG 70039BG 70041BG 70043BG 70044BG 70045BG 70047BG 70048BG 70049BG 70050BG 70051BG 70052BG 70053BG 70054BG 70055BG 70056BG 70059BG 70061BG 70062BG 70065BG 70069BG 70073BG 70075BG 70076BG 70079BG 70082BG 70083BG 70086BG 70087BG 70089BG 70091BG 70092BG 70093BG 70095BG 70096BG 70097BG 70098BG 70101BG 70103BG 70104BG 70106BG 70107BG 70108BG 70109BG 70110BG 70111BG 70112BG 70113BG 70115BG 70116BG 70117BG 70121BG 70124BG 70125BG 70126BG 70129BG 70130BG 70132BG 70133BG 70134BG 70136BG 70137BG 70138BG 70140BG 70142BG 70143BG 70146BG 70147BG 70150BG 70154BG 70156BG 70162BG 70163BG 70164BG 70165BG 70168BG 70169BG 70172BG 70173BG 70179BG 70180BG 70182BG 70184BG 70185BG 70186BG 70187BG 70190BG 70192BG 70193BG 70195BG 70196BG 70198BG 70199BG 70201BG 70202BG 70210BG 70211BG 70212BG 70213BG 70214BG 70215BG 70216BG 70217BG 70218BG 70222BG 70223BG 70224BG 70226BG 70227BG 70229BG 70231BG 70232BG 70233BG 70234BG 70239BG 70242BG 70243BG 70244BG 70247BG 70248BG 70249BG 70250BG 70251BG 70252BG 70253BG 70261BG 70262BG 70267BG 70268BG 70269BG 70270BG 70273BG 70276BG 70277BG 70280BG 70281BG 70282BG 70284BG 70285BG 70287BG 70290BG 70293BG 70295BG 70296BG 70298BG 70300BG 70301BG 70302BG 70304BG 70305BG 70306BG 70308BG 70309BG 70311BG 70313BG 70316BG 70317BG 70318BG 70319BG 70320BG 70321BG 70325BG 70326BG 70327BG 70328BG 70329BG 70330BG 70331BG 70332BG 70333BG 70334BG 70340BG 70341BG 70342BG 70343BG 70346BG 70350BG 70354BG 70355BG 70360BG 70364BG 70366BG 70367BG 70369BG 70372BG 70374BG 70376BG 70378BG 70379BG 70380BG 70381BG 70382BG 70383BG 70384BG 70385BG 70387BG 70390BG 70391BG 70392BG 70393BG 70394BG 70395BG 70397BG 70398BG 70399BG 70400BG 70401BG 70402BG 70403BG 70404BG 70405BG 70406BG 70407BG 70408BG 70409BG 70411BG 70412BG 70413BG 70414BG 70415BG 70418BG 70419BG 70420BG 70421BG 70424BG 70426BG 70427BG 70428BG 70429BG 70430BG 70432BG 70436BG 70437BG 70438BG 70439BG 70443BG 70448BG 70451BG 70453BG 70455BG 70456BG 70459BG 70461BG 70462BG 70464BG 70465BG 70467BG 70468BG 70469BG 70470BG 70471BG 70472BG 70473BG 70475BG 70476BG 70477BG 70479BG 70480BG 70481BG 70482BG 70484BG 70485BG 70486BG 70490BG 70491BG 70493BG 70494BG 70497BG 70500BG 70501BG 70503BG 70504BG 70505BG 70506BG 70507BG 70508BG 70509BG 70511BG 70512BG 70513BG 70514BG 70517BG 70519BG 70521BG 70522BG 70523BG 70525BG 70527BG 70529BG 70530BG 70534BG 70543BG 70544BG 70546BG 70547BG 70548BG 70554BG 70555BG 70556BG 70557BG 70558BG 70559BG 70560BG 70562BG 70563BG 70566BG 70568BG 70569BG 70570BG 70571BG 70572BG 70574BG 70575BG 70576BG 70580BG 70581BG 70582BG 70583BG 70584BG 70585BG 70586BG 70587BG 70588BG 70589BG 70590BG 70591BG 70592BG 70595BG 70596BG 70597BG 70598BG 70600BG 70601BG 70602BG 70604BG 70605BG 70610BG 70611BG 70613BG 70617BG 70618BG 70622BG 70624BG 70626BG 70627BG 70630BG 70633BG 70634BG 70636BG 70637BG 70638BG 70639BG 70640BG 70641BG 70642BG 70644BG 70645BG 70646BG 70647BG 70648BG 70649BG 70650BG 70651BG 70655BG 70657BG 70658BG 70659BG 70660BG 70661BG 70662BG 70664BG 70665BG 70667BG 70669BG 70670BG 70671BG 70672BG 70673BG 70675BG 70676BG 70677BG 70680BG 70681BG 70683BG 70684BG 70686BG 70687BG 70688BG 70690BG 70691BG 70693BG 70694BG 70695BG 70696BG 70697BG 70699BG 70701BG 70702BG 70704BG 70706BG 70708BG 70709BG 70710BG 70711BG 70712BG 70715BG 70717BG 70718BG 70720BG 70721BG 70723BG 70726BG 70728BG 70729BG 70731BG 70735BG 70736BG 70738BG 70739BG 70741BG 70742BG 70743BG 70746BG 70747BG 70748BG 70749BG 70750BG 70751BG 70753BG 70756BG 70757BG 70759BG 70760BG 70761BG 70762BG 70763BG 70764BG 70766BG 70768BG 70769BG 70770BG 70771BG 70772BG 70775BG 70776BG 70777BG 70778BG 70779BG 70780BG 70781BG 70782BG 70783BG 70784BG 70787BG 70791BG 70792BG 70793BG 70795BG 70800BG 70801BG 70802BG 70803BG 70805BG 70806BG 70807BG 70808BG 70810BG 70811BG 70812BG 70813BG 70815BG 70818BG 70819BG 70821BG 70822BG 70823BG 70824BG 70825BG 70826BG 70828BG 70829BG 70830BG 70832BG 70836BG 70841BG 70843BG 70844BG 70848BG 70849BG 70850BG 70851BG 70853BG 70854BG 70855BG 70856BG 70857BG 70858BG 70860BG 70862BG 70864BG 70865BG 70866BG 70867BG 70868BG 70869BG 70871BG 70874BG 70875BG 70876BG 70877BG 70878BG 70879BG 70880BG 70884BG 70886BG 70887BG 70891BG 70892BG 70893BG 70894BG 70895BG 70896BG 70898BG 70899BG 70900BG 70902BG 70904BG 70905BG 70906BG 70907BG 70908BG 70910BG 70913BG 70914BG 70917BG 70918BG 70919BG 70920BG 70921BG 70922BG 70923BG 70924BG 70926BG 70927BG 70928BG 70929BG 70933BG 70935BG 70939BG 70943BG 70944BG 70945BG 70946BG 70947BG 70948BG 70949BG 70950BG 70951BG 70953BG 70954BG 70955BG 70956BG 70958BG 70959BG 70960BG 70962BG 70963BG 70965BG 70966BG 70967BG 70969BG 70970BG 70971BG 70972BG 70974BG 70976BG 70979BG 70980BG 70982BG 70983BG 70984BG 70985BG 70987BG 70988BG 70989BG 70991BG 70992BG 70993BG 70995BG 70996BG 70997BG 70998BG 71000BG 71002BG 71006BG 71008BG 71009BG 71014BG 71015BG 71017BG 71019BG 71021BG 71022BG 71023BG 71025BG 71026BG 71027BG 71031BG 71033BG 71034BG 71036BG 71037BG 71038BG 71041BG 71046BG 71047BG 71048BG 71049BG 71054BG 71055BG 71056BG 71058BG 71059BG 71060BG 71061BG 71063BG 71064BG 71068BG 71069BG 71070BG 71071BG 71072BG 71073BG 71074BG 71075BG 71076BG 71077BG 71078BG 71081BG 71082BG 71087BG 71089BG 71090BG 71092BG 71094BG 71096BG 71097BG 71098BG 71099BG 71100BG 71101BG 71102BG 71103BG 71104BG 71105BG 71107BG 71108BG 71109BG 71110BG 71111BG 71114BG 71115BG 71117BG 71118BG 71119BG 71121BG 71122BG 71123BG 71124BG 71125BG 71126BG 71128BG 71129BG 71130BG 71131BG 71132BG 71133BG 71135BG 71136BG 71137BG 71139BG 71140BG 71147BG 71149BG 71150BG 71151BG 71152BG 71153BG 71154BG 71155BG 71156BG 71157BG 71158BG 71159BG 71160BG 71162BG 71164BG 71165BG 71166BG 71167BG 71168BG 71171BG 71172BG 71173BG 71174BG 71175BG 71176BG 71177BG 71180BG 71181BG 71182BG 71184BG 71186BG 71187BG 71188BG 71191BG 71192BG 71194BG 71195BG 71196BG 71197BG 71198BG 71199BG 71200BG 71208BG 71211BG 71213BG 71215BG 71216BG 71220BG 71222BG 71223BG 71224BG 71225BG 71226BG 71227BG 71229BG 71230BG 71235BG 71236BG 71238BG 71239BG 71240BG 71241BG 71245BG 71247BG 71248BG 71249BG 71250BG 71251BG 71253BG 71256BG 71258BG 71259BG 71260BG 71261BG 71262BG 71263BG 71264BG 71265BG 71266BG 71267BG 71268BG 71269BG 71270BG 71271BG 71272BG 71274BG 71275BG 71276BG 71277BG 71278BG 71282BG 71284BG 71285BG 71288BG 71290BG 71291BG 71292BG 71293BG 71294BG 71295BG 71296BG 71297BG 71298BG 71300BG 71301BG 71302BG 71304BG 71305BG 71309BG 71314BG 71315BG 71316BG 71317BG 71319BG 71320BG 71321BG 71324BG 71328BG 71329BG 71330BG 71332BG 71333BG 71334BG 71336BG 71344BG 71345BG 71346BG 71348BG 71349BG 71350BG 71351BG 71352BG 71353BG 71354BG 71356BG 71360BG 71361BG 71363BG 71364BG 71365BG 71366BG 71367BG 71368BG 71377BG 71381BG 71385BG 71386BG 71390BG 71392BG 71393BG 71394BG 71396BG 71399BG 71400BG 71401BG 71408BG 71411BG 71412BG 71413BG 71414BG 71415BG 71416BG 71417BG 71418BG 71419BG 71420BG 71421BG 71422BG 71423BG 71424BG 71425BG 71426BG 71427BG 71428BG 71429BG 71430BG 71433BG 71434BG 71436BG 71438BG 71439BG 71440BG 71441BG 71442BG 71445BG 71446BG 71447BG 71448BG 71449BG 71450BG 71451BG 71452BG 71453BG 71454BG 71455BG 71456BG 71458BG 71459BG 71461BG 71462BG 71463BG 71464BG 71467BG 71471BG 71472BG 71473BG 71474BG 71475BG 71478BG 71479BG 71480BG 71482BG 71483BG 71484BG 71485BG 71486BG 71488BG 71490BG 71491BG 71492BG 71494BG 71495BG 71497BG 71498BG 71499BG 71500BG 71502BG 71507BG 71508BG 71509BG 71510BG 71512BG 71513BG 71514BG 71515BG 71516BG 71518BG 71519BG 71520BG 71522BG 71524BG 71525BG 71529BG 71530BG 71532BG 71534BG 71535BG 71537BG 71538BG 71539BG 71540BG 71541BG 71544BG 71545BG 71547BG 71548BG 71549BG 71550BG 71552BG 71553BG 71554BG 71555BG 71557BG 71558BG 71560BG 71561BG 71562BG 71563BG 71565BG 71567BG 71568BG 71570BG 71571BG 71572BG 71573BG 71575BG 71576BG 71577BG 71580BG 71581BG 71582BG 71583BG 71584BG 71585BG 71586BG 71587BG 71588BG 71589BG 71590BG 71593BG 71597BG 71598BG 71600BG 71602BG 71605BG 71606BG 71610BG 71611BG 71614BG 71616BG 71617BG 71618BG 71620BG 71621BG 71623BG 71624BG 71627BG 71628BG 71630BG 71632BG 71634BG 71635BG 71636BG 71642BG 71646BG 71647BG 71648BG 71649BG 71650BG 71653BG 71654BG 71655BG 71658BG 71662BG 71663BG 71664BG 71668BG 71669BG 71670BG 71671BG 71672BG 71673BG 71674BG 71675BG 71676BG 71677BG 71678BG 71679BG 71680BG 71681BG 71682BG 71683BG 71684BG 71685BG 71687BG 71689BG 71690BG 71691BG 71692BG 71693BG 71694BG 71695BG 71696BG 71697BG 71699BG 71702BG 71703BG 71704BG 71705BG 71706BG 71711BG 71712BG 71713BG 71714BG 71715BG 71716BG 71717BG 71718BG 71719BG 71720BG 71721BG 71722BG 71726BG 71727BG 71728BG 71729BG 71731BG 71733BG 71735BG 71736BG 71737BG 71739BG 71743BG 71745BG 71746BG 71747BG 71749BG 71750BG 71751BG 71752BG 71753BG 71754BG 71755BG 71756BG 71759BG 71760BG 71761BG 71762BG 71763BG 71764BG 71765BG 71767BG 71768BG 71769BG 71770BG 71771BG 71772BG 71773BG 71774BG 71775BG 71777BG 71778BG 71779BG 71784BG 71785BG 71787BG 71788BG 71789BG 71790BG 71792BG 71795BG 71797BG 71798BG 71799BG 71800BG 71801BG 71802BG 71803BG 71805BG 71807BG 71808BG 71809BG 71810BG 71812BG 71813BG 71815BG 71816BG 71817BG 71818BG 71819BG 71820BG 71821BG 71822BG 71823BG 71824BG 71825BG 71826BG 71829BG 71830BG 71831BG 71832BG 71833BG 71838BG 71840BG 71842BG 71843BG 71844BG 71845BG 71846BG 71847BG 71848BG 71849BG 71851BG 71857BG 71858BG 71859BG 71861BG 71862BG 71863BG 71865BG 71866BG 71880BG 71881BG 71892BG 71893BG 71900BG 71903BG 71907BG 71908BG 71910BG 71911BG 71912BG 71914BG 71915BG 71916BG 71917BG 71918BG 71919BG 71920BG 71921BG 71922BG 71923BG 71924BG 71925BG 71926BG 71927BG 71928BG 71929BG 71932BG 71933BG 71934BG 71935BG 71936BG 71937BG 71938BG 71939BG 71940BG 71941BG 71942BG 71943BG 71944BG 71945BG 71946BG 71947BG 71948BG 71950BG 71952BG 71953BG 71954BG 71956BG 71957BG 71958BG 71959BG 71961BG 71962BG 71963BG 71964BG 71968BG 71970BG 71971BG 71972BG 71975BG 71976BG 71977BG 71978BG 71979BG 71980BG 71981BG 71982BG 71983BG 71984BG 71985BG 71986BG 71987BG 71988BG 71989BG 71990BG 71991BG 71992BG 71993BG 71994BG 71995BG 71996BG 71997BG 71999BG 72000BG 72002BG 72004BG 72005BG 72006BG 72008BG 72009BG 72011BG 72013BG 72016BG 72017BG 72020BG 72025BG 72026BG 72027BG 72028BG 72029BG 72030BG 72032BG 72034BG 72035BG 72036BG 72037BG 72038BG 72040BG 72041BG 72042BG 72043BG 72044BG 72047BG 72048BG 72049BG 72050BG 72051BG 72055BG 72056BG 72058BG 72059BG 72065BG 72072BG 72074BG 72075BG 72076BG 72077BG 72078BG 72079BG 72080BG 72081BG 72084BG 72085BG 72089BG 72091BG 72092BG 72093BG 72094BG 72095BG 72097BG 72098BG 72099BG 72100BG 72101BG 72103BG 72104BG 72105BG 72106BG 72107BG 72108BG 72109BG 72110BG 72111BG 72112BG 72113BG 72114BG 72116BG 72117BG 72118BG 72119BG 72120BG 72121BG 72122BG 72123BG 72124BG 72125BG 72126BG 72127BG 72128BG 72129BG 72130BG 72132BG 72133BG 72134BG 72135BG 72136BG 72137BG 72138BG 72140BG 72141BG 72142BG 72144BG 72148BG 72153BG 72165BG 72166BG 72167BG 72169BG 72173BG 72174BG 72175BG 72176BG 72177BG 72178BG 72183BG 72185BG 72186BG 72187BG 72188BG 72189BG 72190BG 72193BG 72194BG 72195BG 72196BG 72197BG 72198BG 72199BG 72205BG 72206BG 72208BG 72211BG 72214BG 72216BG 72217BG 72221BG 72223BG 72225BG 72236BG 72237BG 72238BG 72239BG 72240BG 72241BG 72242BG 72244BG 72245BG 72246BG 72247BG 72248BG 72250BG 72252BG 72256BG 72258BG 72259BG 72261BG 72262BG 72263BG 72264BG 72268BG 72270BG 72271BG 72272BG 72274BG 72275BG 72277BG 72278BG 72279BG 72280BG 72283BG 72286BG 72291BG 72292BG 72293BG 72294BG 72295BG 72296BG 72297BG 72298BG 72299BG 72301BG 72302BG 72303BG 72304BG 72305BG 72307BG 72308BG 72309BG 72310BG 72311BG 72314BG 72318BG 72320BG 72321BG 72323BG 72324BG 72325BG 72326BG 72328BG 72330BG 72331BG 72332BG 72336BG 72340BG 72341BG 72342BG 72343BG 72347BG 72348BG 72353BG 72354BG 72356BG 72357BG 72358BG 72359BG 72360BG 72361BG 72362BG 72363BG 72364BG 72368BG 72369BG 72370BG 72372BG 72374BG 72376BG 72377BG 72379BG 72380BG 72381BG 72382BG 72383BG 72384BG 72385BG 72386BG 72387BG 72388BG 72389BG 72390BG 72391BG 72394BG 72395BG 72396BG 72398BG 72399BG 72400BG 72402BG 72405BG 72407BG 72417BG 72418BG 72419BG 72420BG 72421BG 72422BG 72423BG 72424BG 72425BG 72427BG 72429BG 72430BG 72431BG 72432BG 72433BG 72448BG"
    }
  ]
}