{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Murrysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98540",
      "recalling_firm": "Philips Respironics, Inc.",
      "address_1": "1001 Murry Ridge Ln",
      "address_2": "",
      "postal_code": "15668-8517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US Nationwide, Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Georgia, French Guiana, Gibraltar, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jersey, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Lithuania, Morocco, Montenegro, North Macedonia, Myanmar, Martinique, Malta, Mexico, Malaysia, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Vietnam, Mayotte, South Africa, Zimbabwe.",
      "recall_number": "Z-1648-2026",
      "product_description": "Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.",
      "product_quantity": "113,717 units",
      "reason_for_recall": "In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.",
      "recall_initiation_date": "20260302",
      "center_classification_date": "20260402",
      "report_date": "20260408",
      "code_info": "All Model No.; All UDI; All Serial No.",
      "more_code_info": ""
    }
  ]
}