{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96493",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.",
      "recall_number": "Z-1647-2025",
      "product_description": "Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, -3 Neck, Model Number 802802201; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-  Porous, Uncemented",
      "product_quantity": "167 units",
      "reason_for_recall": "The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.",
      "recall_initiation_date": "20250324",
      "center_classification_date": "20250421",
      "report_date": "20250430",
      "code_info": "GTIN 00887868376344, All lots distributed with IFU 87-6204-051-99 Rev D or prior"
    }
  ]
}