{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71117",
      "recalling_firm": "Philips Medical Systems, Inc.",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of MO, FL, OH",
      "recall_number": "Z-1643-2015",
      "product_description": "Pinnacle3  Software Version 10.0, Model 459800200841.    Product Usage:  The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data into the system, use that data to construct a plan for radiation therapy and evaluate the plan. Optionally, the qualified medical personnel may output the plan in an electronic or printed form for use by other systems in the delivery of treatment to a patient.",
      "product_quantity": "4",
      "reason_for_recall": "A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users.      Specifically, the dose engine is being passed the wrong snout position.  The snout position is used to determine the penumbra of the beam which includes the calculation of the source si",
      "recall_initiation_date": "20150423",
      "center_classification_date": "20150521",
      "termination_date": "20161005",
      "report_date": "20150527",
      "code_info": "\"System Code S/N\"    1439NML05A, 1117FMM025, DC090775040, 1117FMM03J."
    }
  ]
}