  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98458",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Florida",
      "recall_number": "Z-1637-2026",
      "product_description": "Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937.  Accessory to Baxter Volara Respiratory Therapy system.",
      "product_quantity": "19 units",
      "reason_for_recall": "Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment.  This correction is providing the home caregiver specific instructions for use.",
      "recall_initiation_date": "20260218",
      "center_classification_date": "20260401",
      "report_date": "20260408",
      "code_info": "UDI/DI number 00887761984622",
      "more_code_info": ""
    }
  ]
}