  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65301",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.",
      "recall_number": "Z-1621-2013",
      "product_description": "Bigliani/Flatow¿ The Complete Solution Fukuda Retractors  Small Nonsterile 00-4305-019-30.    Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.",
      "product_quantity": "722 units (includes 2010 market withdrawal)",
      "reason_for_recall": "This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included.  Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.",
      "recall_initiation_date": "20130523",
      "center_classification_date": "20130701",
      "termination_date": "20141218",
      "report_date": "20130710",
      "code_info": "Item Number 00-4305-019-30 with Lot Numbers 61557761, 61591732, 61612275, 61614764, and 77000186.  2010 Market Withdrawal part numbers: Item #:430501930, Lot#s: 60795817, 60834923, 60916795, 60997103, 61017041, 61038697, 61075510, 61128521, 61186400, 61199809, 61253702, 61392455, 61430126"
    }
  ]
}