  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Richmond",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90336",
      "recalling_firm": "Mckesson Medical-Surgical Inc. Corporate Office",
      "address_1": "9954 Maryland Drive",
      "address_2": "Deep Run Iii Ste. 4000",
      "postal_code": "23233",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-1620-2022",
      "product_description": "a. METREX RESEARCH CORP \tCaviCide1\tDISINFECTANT, CLNR CAVICIDE1 1GL (4/CS) METREX,  Model Number: 13-5000.    b.  METREX RESEARCH CORP CaviCide DISINFECTANT, CAVICIDE 8OZ SPRAY MX-1008 METREX Model Number: 13-1008.    c.  METREX RESEARCH CORP CaviCide DISINFECTANT, CAVICIDE GL MX-1000 METREX Catalog #13-1000.    d.  METREX RESEARCH CORP CaviCide DISINFECTANT, CAVICIDE PUMP SPRAY 24OZ METREX Catalog #13-1024.    e. CaviCide1 DISINFECTANT, CLNR CAVICIDE1 24OZ (12/CS) METREX Catalog #13-5024.    f. METREX RESEARCH CORP DISINFECTANT, ENVIROCIDE 24OZ(12/CS) METREX Model Number: 13-3324.    g.  METREX RESEARCH CORP Envirocide DISINFECTANT, ENVIROCIDE GL(4/CS)MX-3300 METREX Model Number: 13-3300",
      "product_quantity": "50cs b. 38 c. 519gal d. 1671  e. 43.25cs f. 33btl g. unk",
      "reason_for_recall": "Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.",
      "recall_initiation_date": "20220525",
      "center_classification_date": "20220825",
      "report_date": "20220831",
      "code_info": "GTIN:   a.  00615375005446; H97113500020.   b.  00615375005309; H97113100824.   c.  00615375005385; H9711310002.   d.  00615375005361, H97113102422.   e.  00615375005415, H97113502426.   f.  00615375005354; H97113332427.   g.  00615375005378; H113330021",
      "more_code_info": ""
    }
  ]
}