{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84780",
      "recalling_firm": "Stryker Neurovascular",
      "address_1": "47900 Bayside Pkwy",
      "address_2": "N/A",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of PA, FL, CA, IL, WI, AZ, AR, FL, TN, MI, MA, CA, NC, WV, DE, OH, NJ, MO, NY, SC, TX, AL, VA, UT, MD, GA, LA, CO, WA, KY, OK, RI, SD, MS, NE,IA, KS, DC.    No US Govt. No Foreign countries.",
      "recall_number": "Z-1620-2020",
      "product_description": "Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-115, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459210(17)211129(10)00000-00    Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-125, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459227(17)211129(10)00000-00    Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-132, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459234(17)211129(10)00000-00  Product Usage: The AXS Vecta Aspiration System, including the AXS Vecta Aspiration Catheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral-M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.",
      "product_quantity": "1405 total products",
      "reason_for_recall": "The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.",
      "recall_initiation_date": "20190410",
      "center_classification_date": "20200331",
      "termination_date": "20210225",
      "report_date": "20200408",
      "code_info": "Catalog # INC-11129-115: (Lot #'s 13170-01,13301-01,13494-01,13543-01,13545-01)  Catalog # INC-11129-125: (Lot #'s 13163-01,13302-01,13500-01,13530-01)  Catalog # INC-11129-132: (Lot #'s 13131-01,13156-01,13158-01,13176-01,13200-01,13201-01,13208-01,13212-01,13213-01,13215-01,13222-01,13238-01,13260-01,13266-01,13289-01,13221-01,13410-01,13412-01,13520-01)"
    }
  ]
}