  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79783",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1608-2018",
      "product_description": "MultiDiagnostic-Eleva with Flat Detector 708038    As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.",
      "product_quantity": "184",
      "reason_for_recall": "The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.",
      "recall_initiation_date": "20180326",
      "center_classification_date": "20180503",
      "termination_date": "20200612",
      "report_date": "20180509",
      "code_info": "System Serial #  40  239  123  58  88  172  129  81  136  159  244  238  82  65  249  70  142  188  128  178  117  45  203  79  229  199  126  158  71  60  187  47  197  87  152  105  76  26  124  139  35  15  66  241  150  86  230  22  206  207  222  146  29  151  111  190  59  228  155  4  233  92  30  221  163  201  91  69  44  205  48  18  34  171  110  175  121  174  43  160  185  23  202  209  95  144  50  141  156  708032/14  72  93  98  193  262  189  118  218  153  33  104  53  216  177  116  38  64  196  49  181  37  180  164  240  135  183  17  125  170  54  NL73  113  90  24  6  27  42  73  246  133  84  214  83  192  149  169  62  61  191  195  100  227  236  97  99  101  102  46  219  103  226  173  89  140  2  36  20  80  32  161  5  51  94  12  176  212  179  25  96  39  154  63  182  157  85  200  234  19  132  21  165  13  120  41  10  56  55  3  7  8  106  114  108  213  14  115  107  31  112"
    }
  ]
}