  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82703",
      "recalling_firm": "Maquet Cardiovascular, LLC",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1601-2019",
      "product_description": "Maquet Getinge-BO-TOP 28903 Adult ECC Pack    Material: 709000093",
      "product_quantity": "N/A",
      "reason_for_recall": "Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured",
      "recall_initiation_date": "20190418",
      "center_classification_date": "20190530",
      "termination_date": "20250728",
      "report_date": "20190605",
      "code_info": "Batch Numbers:   3000045728  3000051160  3000058050"
    }
  ]
}