{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Memphis",
      "address_1": "1800 Pyramid Place",
      "reason_for_recall": "Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.",
      "address_2": "",
      "product_quantity": "400 units",
      "code_info": "Lots: M111063AAF, M111064AAN",
      "center_classification_date": "20130625",
      "distribution_pattern": "Worldwide distribution: USA (nationwide including Puerto Rico) and Internationally to Australia, Brazil, Canada, Chile, Hong Kong, India, Mexico, Netherlands, , New South Wales, Panama, and Paraguay.",
      "state": "TN",
      "product_description": "INFUSE(R) Bone Graft, MEDIUM KIT, REF 7510400AUS, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132    orthopaedic",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Sofamor Danek USA Inc",
      "recall_number": "Z-1594-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65297",
      "termination_date": "20140513",
      "more_code_info": "",
      "recall_initiation_date": "20130528",
      "postal_code": "38132",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}