      "status": "Ongoing",
      "city": "Danvers",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98393",
      "recalling_firm": "Abiomed, Inc.",
      "address_1": "22 Cherry Hill Dr",
      "address_2": "",
      "postal_code": "01923-2575",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "International distribution in the country of Australia, Canada, France, Germany, Kuwait.",
      "recall_number": "Z-1591-2026",
      "product_description": "Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes:     1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA.     2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU.     3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK.    4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.     5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU.     6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK.     7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA.     8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU.     9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.",
      "product_quantity": "62 units",
      "reason_for_recall": "Potential that the \"Purge System Blocked\"  alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.",
      "recall_initiation_date": "20260216",
      "center_classification_date": "20260327",
      "report_date": "20260408",
      "code_info": "All products below with software version 8.2.2 through 8.5 are affected.  1. Product Code: 0042-0000-CA; GTIN: 00813502011272; Serial Numbers: IC2035, IC9416.   2. Product Code: 0042-0000-EU; GTIN: 00813502011289; Serial Numbers: IC1677, IC3049, IC3099, IC3102, IC4358, IC4329, IC5774, IC5757, IC6580, IC6581.  3. Product Code: 0042-0000-UK; GTIN: 00813502011296; Serial Numbers: IC2023, IC1910, IC2704, IC3325, IC3312.  4. Product Code: 0042-0000-US; GTIN: 00813502010022; Serial Numbers: IC1140.  5. Product Code: 0042-0010-EU; GTIN: 00813502010961; Serial Numbers: IC5403, IC5404, IC5635, IC5636, IC5666, IC5667, IC5801, IC5904, IC5918, IC5921, IC6421, IC6422, IC8683.   6. Product Code: 0042-0010-UK; GTIN: 00813502010978; Serial Numbers: IC8101, IC8102, IC8103, IC8104, IC8544, IC9446, IC9449, IC11440.   7. Product Code: 0042-0040-CA; GTIN: 00813502012958; Serial Numbers: IC10306, IC10312, IC12604, IC12891, IC12895, IC12892, IC12605, IC12894.   8. Product Code: 0042-0040-EU; GTIN: 00813502012217; Serial Numbers: IC9014, IC9022, IC9429, IC10408, IC10425.   9. Product Code: 0042-0040-AU; GTIN: 00813502012972; Serial Numbers: IC10163, IC10387, IC11546, IC11566, IC11490, IC11544, IC12008, IC12054, IC11472, IC12151.",
      "more_code_info": ""
    }
  ]
}