{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98432",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.",
      "recall_number": "Z-1587-2026",
      "product_description": "Battery Charging Station;  Model: 0998-00-0802;",
      "product_quantity": "28 units (20 US, 8 OUS)",
      "reason_for_recall": "The Battery Charging Station is an optional dual-bay  charger used to charge the Cardiosave IABP Lithium-  Ion batteries when they are not being used to  operate the IABP. Getinge has identified that a  protruding screw in the left battery bay limits full  insertion of certain batteries and prevents proper  charging. The right bay of the Battery Charging  Station is not affected by this issue.",
      "recall_initiation_date": "20260206",
      "center_classification_date": "20260319",
      "report_date": "20260325",
      "code_info": "Model: 0998-00-0802; UDI: 10607567111964; Serial Numbers: 200019347EAI, 200019847EAI, 200019447EAI, 210018625EAI, 210017925EAI, 210017525EAI, 210018425EAI, 210020144EAI, 210020944EAI, 210022544EAI, 210020244EAI, 210022144EAI, 210020044EAI, 210020544EAI, 210021344EAI, 210023244EAI, 210018025EAI, 210019025EAI, 210021144EAI, 210021044EAI, 210022344EAI, 210019225EAI, 210019125EAI, 210018525EAI, 210022844EAI, 200017647EAI, 210019725EAI, 210022744EAI;",
      "more_code_info": ""
    }
  ]
}