{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98411",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: Nationwide Distribution;  International: Australia, Germany, Belgium, France, Spain UK, New Zealand;",
      "recall_number": "Z-1572-2026",
      "product_description": "MICS3 Angled Sagittal Saw Attachment;  Part Number: 210490",
      "product_quantity": "1098 units",
      "reason_for_recall": "A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.",
      "recall_initiation_date": "20260202",
      "center_classification_date": "20260318",
      "report_date": "20260325",
      "code_info": "Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;",
      "more_code_info": ""
    }
  ]
}